Agreement between Fourier-domain and swept-source optical coherence tomography used for optic nerve head measurements.

Kudsieh B, Fernandez-Vigo J I, De-Pablo-Gómez-de-Liaño L, Fernández-Vigo C, Ruiz Moreno J M, Fernández-Vigo J Á

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the correlation between optic nerve head measurements generated by Fourier-domain (FD) and swept-source (SS) optical coherence tomography (OCT) both in healthy Caucasian subjects and patients with primary open angle glaucoma (POAG).

MATERIALS AND

METHODS

This was a cross-sectional study of the right eyes of 118 subjects. In each participant, the measurements of disc area, cup to disc ratio (CDR), vertical cup to disc ratio (VCDR), rim area and rim volume were performed consecutively by FD-OCT and then SS-OCT. Participant age, gender and spherical equivalent were also recorded. Agreement between the two devices was assessed through intraclass correlation coefficients (ICC) and Bland-Altman plots.

RESULTS

The study sample consisted of 95 healthy eyes and 23 eyes with glaucoma. Mean participant age was 48.6±20.0 years, 54.2% were female, and mean spherical equivalent was -1.6±3.0 diopters. FD-OCT and SS-OCT measurements were respectively: mean disc area 1.79±0.3 vs 1.83±0.3 mm2 (ICC=0.71), mean CDR 0.38±0.2 vs 0.33±0.2 (ICC=0.91), mean VCDR 0.58±0.2 vs 0.52±0.2 (ICC=0.92), mean rim area 1.05±0.4mm2 vs 1.03±0.5mm2 (ICC=0.29), and mean rim volume 0.14±0.11 vs 0.21±0.17mm3 (ICC=0.53). Good agreement between the devices was noted for rim area and rim volume in glaucoma subjects (ICC=0.76 and 0.68 respectively), while weak agreement was observed for these variables in healthy subjects (ICC≤0.50).

CONCLUSIONS

The CDR and VCDR measurements provided by FD and SS OCT showed excellent agreement for the overall sample. When the devices were used for rim measurements, agreement was excellent only in the POAG patients.

Для цитирования: Kudsieh B, Fernandez-Vigo J I, De-Pablo-Gómez-de-Liaño L, Fernández-Vigo C, Ruiz Moreno J M, Fernández-Vigo J Á. Agreement between Fourier-domain and swept-source optical coherence tomography used for optic nerve head measurements.. Journal francais d'ophtalmologie. 2019. DOI: 10.1016/j.jfo.2019.06.016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.