Anterior Segment Optical Coherence Tomography Signs of Local Dilatation Effect of a Micro-Stent on Schlemm's Canal.

Gillmann Kevin, Bravetti Giorgio Enrico, Mansouri Kaweh, Mermoud André

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The iStent inject® (Glaukos Corporation, CA, USA) is a relatively new device designed to be implanted ab-interno through the trabecular meshwork. This is, to the best of our knowledge, the first in-vivo description of a trabecular bypass device visualised with anterior segment optical coherence tomography (AS-OCT), and report of its structural effect on Schlemm's canal.

CASE REPORT: A 74 year-old female patient suffering from long-standing primary open-angle glaucoma and nuclear sclerosis underwent cataract surgery combined with the implantation of two iStent injects®. Surgery was uncomplicated and achieved intraocular pressure (-1 mmHg) and medication (-2 molecules) reduction at 6 months. Under AS-OCT (Spectralis OCT, Heidelberg Engineering AG, Germany) the stent appears as a 300 μm long hyper reflective hollow device within the trabecular meshwork. Approximately a third of it protruded into the anterior chamber. Profound OCT signal loss was notable within the shadow of the device. A second AS-OCT section 500 μm beside the microstent shows a markedly dilated Schlemm's canal, with a major diameter of 390 μm.

DISCUSSIONS: This report confirms that AS-OCT is a suitable technique to assess microstent positioning, and provides a first report on the in-vivo appearance of a functioning stent. It also indicates that iStent injects® could have a tangible effect on adjacent portions of Schlemm's canal with, in this case, a 220% increase in canal diameter compared to the observed average (122 μm). This suggests the IOP-lowering effect of trabecular bypass devices could rely on a dual mechanism involving Schlemm's canal dilatation.

Для цитирования: Gillmann Kevin, Bravetti Giorgio Enrico, Mansouri Kaweh, Mermoud André. Anterior Segment Optical Coherence Tomography Signs of Local Dilatation Effect of a Micro-Stent on Schlemm's Canal.. Nepalese journal of ophthalmology : a biannual peer-reviewed academic journal of the Nepal Ophthalmic Society : NEPJOPH. 2018. DOI: 10.3126/nepjoph.v10i2.23030.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.