Peripheral nerve stimulation by time-varying magnetic fields.

Abart J, Eberhardt K, Fischer H, Huk W, Richter E, Schmitt F, Storch T, Tomandl B, Zeitler E

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: A study was performed to assess the stimulation threshold for healthy adults using sinusoidally oscillating gradients.

METHOD: One hundred thirteen healthy adults were examined in the study. ECG and physiological parameters were measured. All measurements were performed of both the head and the abdomen. The subjects were measured in the supine position with the region of interest positioned in the center of the gradient coils. The measurement was performed for three orthogonal, four oblique, and double oblique orientations.

RESULTS

No volunteer reported painful, severe stimulation. The mean thresholds for peripheral stimulation in head and body measurement were similar: 85.5% of stimulation during examination of the head and 87.6% during measurements of the abdomen were reported when the y-gradient was used.

CONCLUSION

The greatest frequency of reported stimulations occurs when the y-gradient is used. This was confirmed by the results and supports the hypothesis that orthogonal to the y-axis the body has its largest conductive loop, resulting in the strongest peripheral stimulation.

Для цитирования: Abart J, Eberhardt K, Fischer H, Huk W, Richter E, Schmitt F, Storch T, Tomandl B, Zeitler E. Peripheral nerve stimulation by time-varying magnetic fields.. Journal of computer assisted tomography. 1997. DOI: 10.1097/00004728-199707000-00005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.