MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/j.jcrs.0000000000000539

Медицинская библиотека · Journal of cataract and refractive surgery · 2021

Wavefront-guided myopic laser in situ keratomileusis with a high-resolution Hartmann-Shack aberrometer and a new nomogram.

Maloney Robert K, Kraff Colman R, Coleman Stephen C

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate visual, refractive, aberrometric, and patient-reported outcomes of wavefront-guided (WFG) myopic laser in situ keratomileusis (LASIK) using a high-resolution Hartmann-Shack aberrometer (iDesign Advanced WaveScan system) with a new nomogram and to determine whether the new nomogram resolved the mild undercorrection that occurs with the manufacturer's default settings.

SETTING

Three private LASIK practices.

DESIGN

Prospective, open-label, noncomparative, multicenter study.

METHODS

One hundred ninety eyes of 95 patients underwent bilateral WFG LASIK for the correction of myopia or myopic astigmatism. A new nomogram was used, which effectively adjusted the wavefront-measured refraction sphere up or down to equal the manifest refraction sphere. Patients were followed up for 6 months.

RESULTS

Eighty-four patients completed the final follow up. At 6 months, 162 (96.4%) of 168 eyes achieved monocular uncorrected distance visual acuity of 20/20 or better. No eye lost 2 or more lines of corrected distance visual acuity. The safety and efficacy indices were 1.12 and 1.09, respectively; 164 (98%) of 168 eyes had manifest refraction spherical equivalent within ±0.50 diopters (D) of emmetropia, and 154 (92%) of 168 eyes had residual manifest refractive astigmatism of 0.50 D or less. Fewer patients experienced burning, stinging, soreness, and irritation postoperatively than preoperatively. Eighty-one (96%) of 84 patients reported improved quality of life.

CONCLUSIONS

WFG myopic LASIK using a high-resolution Hartmann-Shack aberrometer and a new nomogram resolved the undercorrection with the manufacturer's default settings. The treatment was safe and effective with excellent visual and refractive outcomes, high patient satisfaction, and improved quality of life.

Для цитирования: Maloney Robert K, Kraff Colman R, Coleman Stephen C. Wavefront-guided myopic laser in situ keratomileusis with a high-resolution Hartmann-Shack aberrometer and a new nomogram.. Journal of cataract and refractive surgery. 2021. DOI: 10.1097/j.jcrs.0000000000000539.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.