DOI 10.1097/j.jcrs.0000000000000539
Wavefront-guided myopic laser in situ keratomileusis with a high-resolution Hartmann-Shack aberrometer and a new nomogram.
Аннотация
PURPOSE: To evaluate visual, refractive, aberrometric, and patient-reported outcomes of wavefront-guided (WFG) myopic laser in situ keratomileusis (LASIK) using a high-resolution Hartmann-Shack aberrometer (iDesign Advanced WaveScan system) with a new nomogram and to determine whether the new nomogram resolved the mild undercorrection that occurs with the manufacturer's default settings.
SETTING
Three private LASIK practices.
DESIGN
Prospective, open-label, noncomparative, multicenter study.
METHODS
One hundred ninety eyes of 95 patients underwent bilateral WFG LASIK for the correction of myopia or myopic astigmatism. A new nomogram was used, which effectively adjusted the wavefront-measured refraction sphere up or down to equal the manifest refraction sphere. Patients were followed up for 6 months.
RESULTS
Eighty-four patients completed the final follow up. At 6 months, 162 (96.4%) of 168 eyes achieved monocular uncorrected distance visual acuity of 20/20 or better. No eye lost 2 or more lines of corrected distance visual acuity. The safety and efficacy indices were 1.12 and 1.09, respectively; 164 (98%) of 168 eyes had manifest refraction spherical equivalent within ±0.50 diopters (D) of emmetropia, and 154 (92%) of 168 eyes had residual manifest refractive astigmatism of 0.50 D or less. Fewer patients experienced burning, stinging, soreness, and irritation postoperatively than preoperatively. Eighty-one (96%) of 84 patients reported improved quality of life.
CONCLUSIONS
WFG myopic LASIK using a high-resolution Hartmann-Shack aberrometer and a new nomogram resolved the undercorrection with the manufacturer's default settings. The treatment was safe and effective with excellent visual and refractive outcomes, high patient satisfaction, and improved quality of life.