DOI 10.1097/j.jcrs.0000000000000489
Anterior capsule coverage and rotational stability of an acrylic toric intraocular lens.
Аннотация
PURPOSE: To investigate factors related to the rotational stability of an acrylic toric intraocular lens (IOL).
SETTING
Four ophthalmic surgical sites in Japan.
DESIGN
Prospective case series.
METHODS
The study included 120 eyes of 120 patients undergoing phacoemulsification and implantation of a toric IOL (AcrySof IQ, Alcon Laboratories, Inc.). At 1 hour postoperatively, the area of continuous curvilinear capsulorhexis (CCC) was measured, and the state of anterior capsule coverage on the IOL optic (total on or partial on) was recorded. The toric IOL axis orientation was assessed at the end of surgery and at 1 hour, 1 week, 1 month, and 6 months postoperatively. Multiple regression analysis was performed to explore any clinical factors relevant to IOL rotation from the end of surgery to 6 months postoperatively. The explanatory variables included age, anterior chamber depth preoperatively, axial length, type of corneal astigmatism (with-the-rule, against-the-rule, or oblique astigmatism), area of CCC, state of anterior capsule overlap on IOL optic (total coverage vs partial coverage), and surgical sites (surgeons).
RESULTS
The multiple regression analysis in 110 eyes of 110 patients indicated that anterior capsule overlap on the IOL optic was the only variable associated with IOL rotation at 6 months postoperatively (P = .0482). The mean absolute rotation at 6 months was 1.96 ± 1.81 degrees in the total on group and 3.79 ± 3.12 degrees in the partial on group (P = .0004).
CONCLUSIONS
Rotational stability of a single-piece, acrylic toric IOL was better in eyes with total anterior capsule coverage than that in those with partial anterior capsule coverage on the IOL optic.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.