Intrinsic Optical Properties of Boston Keratoprosthesis.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Hui Pui-Chuen, Pereira Leonardo A, Dore Renald, Chen Shengtong, Taniguchi Elise, Chodosh James, Dohlman Claes H, Paschalis Eleftherios I

Аннотация

PURPOSE: To benchmark the optical performance of Boston Keratoprosthesis (B-KPro).

METHODS: Back focal lengths (BFL) of B-KPros for various eye axial lengths were measured using an optical bench, International Organization for Standardization-certified for intraocular lens characterization, and compared against manufacturer's specification. The modulation transfer function (MTF) and the resolution efficiencies were measured. The theoretical geometry-dependent higher-order aberrations (HOA) were calculated. The devices were characterized with optical profilometry for estimating the surface scattering. Aberration correction and subsequent image quality improvement were simulated in CODE-V. Natural scene-imaging was performed in a mock ocular environment. Retrospective analysis of 15 B-KPro recipient eyes were presented to evaluate the possibility of achieving 20/20 best-corrected visual acuity (BCVA).

RESULTS: BFL measurements were in excellent agreement with the manufacturer-reported values (r = 0.999). The MTF specification exceeded what is required for achieving 20/20 visual acuity. Astigmatism and field curvature, correctable in simulations, were the primary aberrations limiting imaging performance. Profilometry of the anterior surface revealed nanoscale roughness (root-mean-square amplitude, 30-50 nm), contributing negligibly to optical scattering. Images of natural scenes obtained with a simulated B-KPro eye demonstrated good central vision, with 10/10 visual acuity (equivalent to 20/20). Full restoration of 20/20 BCVA was obtainable for over 9 years in some patients.

CONCLUSIONS: Theoretical and experimental considerations demonstrate that B-KPro has the optical capacity to restore 20/20 BCVA in patients. Further image quality improvement can be anticipated through correction of HOAs.

TRANSLATIONAL RELEVANCE: We establish an objective benchmark to characterize the optics of the B-KPro and other keratoprosthesis and propose design changes to allow improved vision in B-KPro patients.