Multizonal Design Multifocal Intraocular Lens-Induced Astigmatism According to Orientation.

Eom Youngsub, Yang Seul Ki, Yoon Eun Gyu, Choi Ji Nyeong, Ryu Dongok, Kim Dae Wook, Kim Jun-Heon, Song Jong Suk, Kim Sug-Whan, Kim Hyo Myung

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the differences in intraocular lens (IOL)-induced astigmatism according to differences in orientation of a multizonal multifocal IOL, the Precizon Presbyopic NVA IOL (Ophtec BV).

METHODS

The clinical study reviewed 80 eyes from 40 patients with cataracts who underwent Precizon Presbyopic IOL implantation. Residual astigmatism, as measured by autorefraction and manifest refraction, was investigated using vector analysis of eyes implanted with vertical (90 ± 30 degrees) and horizontal (180 ± 30 degrees) orientations of the first near segment of the IOL. In the ray-tracing simulation study, pseudophakic eyes with the Precizon Presbyopic IOL were modeled. The modulation transfer function (MTF) of each case was compared with respect to the amount of corneal astigmatism of the model eyes and the orientation of the first near segment.

RESULTS

The mean IOL-induced astigmatism measured by autorefraction was 0.68 ± 0.58 diopters (D) at 1 degree in the vertical orientation of the first near segment (n = 52) and 1.05 ± 0.81 D at 96 degrees in the horizontal orientation (n = 28). However, the mean IOL-induced astigmatism measured by manifest refraction was 0.14 ± 0.44 D at 171 degrees and 0.46 ± 0.40 D at 95 degrees. The MTF analysis showed that the highest MTF values were measured in eyes without corneal astigmatism in both the vertically and horizontally implanted IOLs.

CONCLUSIONS

Autorefraction measurement indicates induction of with-the-rule astigmatism by the Precizon Presbyopic IOL when implanted vertically with respect to the first near segment, and against-the-rule astigmatism when implanted horizontally. However, this astigmatism is clinically insignificant according to manifest refraction and ray-tracing simulation. [J Refract Surg. 2020;36(11):740-748.].

Для цитирования: Eom Youngsub, Yang Seul Ki, Yoon Eun Gyu, Choi Ji Nyeong, Ryu Dongok, Kim Dae Wook, Kim Jun-Heon, Song Jong Suk, Kim Sug-Whan, Kim Hyo Myung. Multizonal Design Multifocal Intraocular Lens-Induced Astigmatism According to Orientation.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2020. DOI: 10.3928/1081597x-20200828-02.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.