DOI 10.1097/j.jcrs.0000000000000428
Rethinking the optimal methods for vector analysis of astigmatism.
Аннотация
PURPOSE: To evaluate the accuracy and usefulness of certain methods of analyzing astigmatic vectors.
SETTING
Cullen Eye Institute, Baylor College of Medicine, Houston, Texas.
DESIGN
Case samples.
METHODS
Using 2 sample cases for analysis of corneal surgically induced astigmatism and an actual toric intraocular lens (IOL) case, univariate analyses from the ASSORT program were compared to double-angle plots of preoperative and postoperative astigmatism and prediction errors.
RESULTS
Certain univariate figures for analyzing the 2 sample cases were misleading. For the toric IOL case, some of the key outcome vectors were inaccurate.
CONCLUSIONS
ASSORT's univariate analysis of astigmatic vectors can be unpredictably erroneous and misleading. Recommended vector analyses should include double-angle plots with centroids and confidence ellipses of preoperative and postoperative astigmatism and the prediction errors, along with means and standard deviations of these vector magnitudes.