m+Medelevator.RU
← Поиск публикаций
Journal of cataract and refractive surgery · 2021

Long-term efficacy and safety results after iris-fixated foldable phakic intraocular lens for myopia and astigmatism: 6-year follow-up.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Monteiro Tiago, Correia Fernando Faria, Franqueira Nuno, Mendes José Carlos, Pinto Christophe, Vaz Fernando

Аннотация

PURPOSE: To evaluate the long-term efficacy and safety of iris-fixated foldable phakic intraocular lenses (pIOLs) for the management of myopia and astigmatism after 6-year follow-up.

SETTING: Ophthalmology Department, Hospital Braga, Portugal.

DESIGN: Retrospective cohort study.

METHODS: Patients included underwent iris-fixated foldable Artiflex pIOL implantation between January 2010 and December 2013. Follow-up evaluations were performed 6 months postoperatively and every 12 months until 72 months. Preoperative and follow-up data on uncorrected and corrected distance visual acuity (CDVA), manifest refraction, endothelial cell density (ECD), and anterior chamber depth (ACD) were analyzed.

RESULTS: A total of 177 eyes of 98 patients with a mean age of 32.33 ± 7.13 years were analyzed. There was a statistically significant improvement in CDVA from 0.09 ± 0.02 logMAR preoperatively to 0.04 ± 0.02 logMAR (P < .0001) at 6 years postoperatively. The spherical equivalent demonstrated a statistically significant improvement from -9.50 ± 2.93 diopters (D) to -0.41 ± 0.45 D at 6 years. At 72 months, the efficacy and safety indexes were 0.94 and 1.15, respectively. A mean loss of 31.77 cells/mm2 (1.19%) was observed each year until the sixth year. The overall mean ECD loss after correction for the physiological loss was 3.02% after 6 years. No correlations were observed between the mean ECD loss and the preoperative ACD.

CONCLUSIONS: The Artiflex pIOL was a safe, effective, and a stable option to correct myopia. A mean ECD loss of 3.02% was observed over 6 years. No pIOL was explanted due to corneal decompensation.