DOI 10.1097/j.jcrs.0000000000000395

Медицинская библиотека · Journal of cataract and refractive surgery · 2021

Optimized CorT Total to compare Scheimpflug vs dual Scheimpflug/Placido imaging devices.

Ferguson Tanner J, Piccinini Andre L, Thulasi Praneetha, Ong James K Y, Stamatelatos George, Alpins Noel, Randleman J Bradley

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the manufacturer-provided measures of total corneal power (TCP) generated by Scheimpflug and dual Scheimpflug/Placido imaging compared with corneal topographic astigmatism calculated on the basis of measured TCP data (CorT Total).

SETTING

Emory University, Atlanta, Georgia, USA.

DESIGN

Retrospective case series.

METHODS

TCP values were exported from virgin 209 eyes that underwent imaging with both the Scheimpflug (Pentacam HR) and dual Scheimpflug-Placido (Galilei G4) imaging devices to compute an optimized CorT Total. The standard deviation of the ocular residual astigmatism (ORAsd), which serves as a value describing the vectoral difference between the corneal astigmatism measure and manifest refractive cylinder at the corneal plane, was evaluated for all eyes to compare manufacturer-provided measurements vs the optimized CorT Total.

RESULTS

The Scheimpflug CorT Total had the lowest ORAsd (0.306 diopter [D]; spherical equivalent [SE] 0.018) of all the parameters evaluated, although the difference was not statistically significant (P = .22) from the dual Scheimpflug/Placido CorT Total (0.32 2 D; SE 0.017). For the Scheimpflug device, the CorT Total had a statistically significant lower (P < .05) ORAsd in comparison to the best measure on the device (total corneal refractive power apex zone 2 mm: 0.324 D; SE 0.021). For dual Scheimpflug/Placido measurements, the CorT Total had the lowest ORAsd (0.322 D; SE 0.017), but the difference was not statistically significant (P = .43) from the lowest manufacturer-provided measure (TCP 2).

CONCLUSIONS

CorT Total generated with the Scheimpflug device corresponded better with the manifest refractive cylinder than all measures of total corneal astigmatism calculated by the software from both the Scheimpflug and the dual Scheimpflug/Placido devices.

Для цитирования: Ferguson Tanner J, Piccinini Andre L, Thulasi Praneetha, Ong James K Y, Stamatelatos George, Alpins Noel, Randleman J Bradley. Optimized CorT Total to compare Scheimpflug vs dual Scheimpflug/Placido imaging devices.. Journal of cataract and refractive surgery. 2021. DOI: 10.1097/j.jcrs.0000000000000395.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.