Icare versus Goldmann in a randomised middle-aged population : The influence of central corneal thickness and refractive errors.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Stoor Katri, Karvonen Elina, Ohtonen Pasi, Liinamaa M Johanna, Saarela Ville
Аннотация
PURPOSE: The aim of this study was to compare the measurements of intraocular pressure by two tonometers, the Icare rebound tonometer and the Goldmann applanation tonometer, in a randomised screening study. The influence of refraction and central corneal thickness on the measurements was also evaluated.
METHODS: Intraocular pressure was measured with rebound tonometer and Goldmann applanation tonometer in 1266 participants; refraction and central corneal thickness were also determined. One randomised eye of each participant was selected for this report's analysis. A Bland-Altman plot was used to compare the values obtained with the two devices.
RESULTS: The correlation between rebound tonometer and Goldmann applanation tonometer was good: the intraclass correlation coefficient (r) between the two methods was 0.735 (p < 0.001). The mean difference (rebound tonometer-Goldmann applanation tonometer) was 0.11 ± 2.3 mmHg. The difference was not statistically significant (95% confidence interval: 0.11 to 0.13, p = 0.09). With increasing central corneal thickness, not only did intraocular pressure values with both devices increase, but the difference between them also increased. Refraction (spherical equivalent) did not influence intraocular pressure or the rebound tonometer-Goldmann applanation tonometer difference. However, high astigmatism (≥2D) exerted an influence on intraocular pressure values taken with Goldmann applanation tonometer.
CONCLUSION: Measurements with rebound tonometer and Goldmann applanation tonometer are relatively uniform although rebound tonometer slightly overestimated intraocular pressure. Both rebound tonometer and Goldmann applanation tonometer and the difference between these devices were affected by central corneal thickness but not by refraction. Higher astigmatism affected Goldmann applanation tonometer more than rebound tonometer. It is concluded that rebound tonometer is a reliable method for measuring intraocular pressure in a population-based screening study.
METHODS: Intraocular pressure was measured with rebound tonometer and Goldmann applanation tonometer in 1266 participants; refraction and central corneal thickness were also determined. One randomised eye of each participant was selected for this report's analysis. A Bland-Altman plot was used to compare the values obtained with the two devices.
RESULTS: The correlation between rebound tonometer and Goldmann applanation tonometer was good: the intraclass correlation coefficient (r) between the two methods was 0.735 (p < 0.001). The mean difference (rebound tonometer-Goldmann applanation tonometer) was 0.11 ± 2.3 mmHg. The difference was not statistically significant (95% confidence interval: 0.11 to 0.13, p = 0.09). With increasing central corneal thickness, not only did intraocular pressure values with both devices increase, but the difference between them also increased. Refraction (spherical equivalent) did not influence intraocular pressure or the rebound tonometer-Goldmann applanation tonometer difference. However, high astigmatism (≥2D) exerted an influence on intraocular pressure values taken with Goldmann applanation tonometer.
CONCLUSION: Measurements with rebound tonometer and Goldmann applanation tonometer are relatively uniform although rebound tonometer slightly overestimated intraocular pressure. Both rebound tonometer and Goldmann applanation tonometer and the difference between these devices were affected by central corneal thickness but not by refraction. Higher astigmatism affected Goldmann applanation tonometer more than rebound tonometer. It is concluded that rebound tonometer is a reliable method for measuring intraocular pressure in a population-based screening study.