OBJECTIVE

The Contour Neurovascular System (CNS) is a novel intrasaccular flow-disrupting device for the treatment of intracranial aneurysms. This study presents the authors' institutional experience and midterm follow-up results with this device.

METHODS

Seventy-six consecutive patients (mean ± SD age 58.9 ± 12.4 years) were treated with the CNS for 76 aneurysms (63 unruptured, 10 recurrent, and 3 ruptured). Aneurysm characteristics, procedural details, and clinical/angiographic outcomes were retrospectively evaluated.

RESULTS

The most common aneurysm locations were the basilar tip in 25 (32.9%) cases and the anterior communicating artery in 22 (28.9%). There were 18 (23.7%) sidewall aneurysms. The mean ± SD aneurysm width was 5.6 ± 2.5 mm, and 68 (89.5%) aneurysms were wide-necked. The CNS was successfully implanted in 68 (89.5%) aneurysms, with 11 cases of additional coiling and 1 case each of additional stent and balloon. There were 3 (3.9%) thromboembolic events, of which 1 (1.3%) was symptomatic (a major ischemic stroke) and resulted in morbidity. There were no hemorrhagic events or deaths. At last available follow-up (mean 12 months), 32/56 (57.1%) aneurysms were completely occluded, 16/56 (28.6%) had neck remnants, and 8/56 (14.3%) had an aneurysm remnant. Three (5.4%) aneurysms were retreated.

CONCLUSIONS

The use of the CNS was safe and feasible but proper aneurysm selection appears to be required, which may be facilitated with increasing operator experience and further study of this device. Similar to other intrasaccular flow disrupters, midterm complete occlusion rates are moderate but may increase with longer follow-up.

Для цитирования: Radomi Alexandra, Goertz Lukas, Forbrig Robert, Pflaeging Muriel, Kabbasch Christoph, Liebig Thomas. Safety and efficacy of the Contour Neurovascular System for intracranial aneurysm treatment: a retrospective analysis of 76 patients.. Journal of neurosurgery. 2024. DOI: 10.3171/2024.5.jns232852.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.