DOI 10.1016/j.jcrs.2006.02.055

Медицинская библиотека · Journal of cataract and refractive surgery · 2006

Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens.

Muñoz Gonzalo, Albarrán-Diego César, Montés-Micó Robert, Rodríguez-Galietero Antonio, Alió Jorge L

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To determine whether implantation of an intraocular lens (IOL) with a modified prolate anterior surface (Tecnis Z9000, AMO) results in reduced spherical aberration and improved contrast sensitivity after cataract surgery.

SETTING

Hospital NISA Virgen del Consuelo, Valencia, Spain.

METHODS

In an intraindividual randomized prospective study of 30 patients with bilateral cataract, the Tecnis Z9000 IOL was compared with 2 IOLs with spherical surfaces, the AR40e (AMO) and the Stabibag (Ioltech). Ocular aberrations for a 4.0 mm pupil and 6.0 mm pupil were measured with a Hartmann-Shack aberrometer. Quality of vision was measured by visual acuity and contrast sensitivity under mesopic and photopic conditions.

RESULTS

Eyes with the Tecnis Z9000 IOL had significantly less spherical aberration and a greater Strehl ratio after surgery. Significant reduction in coma aberration was also found in some cases. Refraction, visual acuity, and contrast sensitivity were not significantly different between the 3 IOL groups.

CONCLUSION

A significant reduction in spherical aberration after Tecnis Z9000 IOL implantation was achieved, but visual acuity and contrast sensitivity were not affected by the aspheric silicone IOL compared to spherical acrylic IOLs.

Для цитирования: Muñoz Gonzalo, Albarrán-Diego César, Montés-Micó Robert, Rodríguez-Galietero Antonio, Alió Jorge L. Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens.. Journal of cataract and refractive surgery. 2006. DOI: 10.1016/j.jcrs.2006.02.055.