DOI 10.1007/s00066-025-02451-7
Efficient workflow and clinical validation of in-house 3D-printed vaginal cylinders for high-dose-rate brachytherapy for gynecologic malignancies.
Аннотация
PURPOSE: This study developed an efficient methodology for in-house 3D-printed vaginal cylinders for gynecologic tumor treatment by evaluating their radiation attenuation, geometric accuracy, and efficacy. Ultimately, we aim to establish a simple, cost-effective approach that facilitates broad clinical adoption.
METHODS
Patient-specific vaginal cylinders were designed based on anatomical contours from the treatment planning system (TPS) using CAD software. The process was optimized to minimize manpower and time costs. Radiation attenuation of the 3D printer material was compared with that of water using the Monte Carlo method. Geometric accuracy was automatically analyzed via an in-house MATLAB program. Efficacy was assessed in cases of postoperative vaginal stump recurrence and vaginal cancer with paravaginal invasion.
RESULTS
The tumor shape, delineated by TPS, was imported into CAD software, and the catheter pathway model, designed via subtraction processing, was placed at the optimal position and angle. The design process took approximately 15 min, and the entire workflow was completed within a week, demonstrating its practicality for clinical use. The radiation attenuation error was < 3% compared with water, and the geometric accuracy error was < 0.2 mm. The patient-specific vaginal cylinder provided a favorable dose distribution and was effective in complex cases.
CONCLUSION
A feasible workflow was established, allowing in-house design and manufacturing with reduced manpower and time costs. With no material or processing issues, this approach is safe, practical, and promising for widespread adoption in personalized brachytherapy.