DOI 10.1093/europace/eum021

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2007

The contribution of rate adaptive pacing with single or dual sensors to health-related quality of life.

van Hemel Norbert M, Holwerda Klaas J, Slegers Paul C, Spierenburg Han A M, Timmermans Alphons A J M, Meeder Joan G, van der Kemp Peter, Kelder Johannes C, Stofmeel Monique A M, Sensor and Quality of Life (SQL) Investigators

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

AIMS: The characteristics of sensors to perform rate adaptive pacing are well established but whether their contribution improves health-related quality of life (QoL) remains disputable. To compare the effects on QoL with an integrated dual sensor [minute ventilation (MV) and acceleration, TT sensor] with a single MV sensor, and with no rate adaptive pacing.

METHODS AND

RESULTS

This Dutch multi centre, prospective, single- (patient) blind study was performed in patients after first pacemaker (PM) implant for sick sinus syndrome or AV block. After a 3-month 'sensor off'-period following DDD PM implantation, where the latter 2 months permitted the MV sensor to learn the intrinsic rhythm, a 2-month period of DDDR with TT sensor or 2 months of DDDR with MV sensor, subsequently the two modes were crossed over. Quality of life was determined with Aquarel, the disease-specific instrument for PM patients. Heart rate, percentages of sensor driven and intrinsic rhythm were retrieved from PM memories. Sixty-four patients completed the 7-month study. In sick sinus patients, percentages of sensor-driven pacing occurred significantly more frequently than in AV block patients After implant QoL improved significantly: before 71.3 and after 83.5% (P < 0.001) measured with Aquarel and in 3 of 9 SF-36 scales, but no significant additive QoL benefit with dual or MV sensor pacing was observed. Pacing diagnosis, percentages of rate adaptive pacing, and heart rate influencing medication did not influence this result.

CONCLUSION

Pacemaker implantation strongly improves QoL, but neither single- nor dual- sensor-driven pacing offered additional improvement in QoL during the initial 8 months after the first PM implant.

Для цитирования: van Hemel Norbert M, Holwerda Klaas J, Slegers Paul C, Spierenburg Han A M, Timmermans Alphons A J M, Meeder Joan G, van der Kemp Peter, Kelder Johannes C, Stofmeel Monique A M, Sensor and Quality of Life (SQL) Investigators. The contribution of rate adaptive pacing with single or dual sensors to health-related quality of life.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2007. DOI: 10.1093/europace/eum021.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.