Is a dual-sensor pacemaker appropriate in patients with sino-atrial disease? Results from the DUSISLOG study.

Padeletti Luigi, Pieragnoli Paolo, Di Biase Luigi, Colella Andrea, Landolina Maurizio, Moro Eugenio, Orazi Serafino, Vicentini Alfredo, Maglia Giampiero, Pensabene Orazio, Raciti Giovanni, Barold S Serge

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Rate-responsive pacemakers (PMs) are often supplied with accelerometer (XL) and minute ventilation (MV) sensors to provide a physiologic rate response according to patient needs. No information is available about the real benefit of dual-sensor rate-responsive pacing on the daily life of patients.

METHODS

DUSISLOG (Dual Sensor vs Single Sensor comparison using patient activity LOGbook) is a two-arm prospective, randomized, multicenter study that enrolled 105 patients who received a rate-responsive PM (Insignia), Guidant Corp.). After 1 month of DDD pacing at 60 ppm lower rate, a single sensor (XL or MV, randomized) was activated for 3 months at the manufacturer's suggested nominal settings, followed by a 3-month period with dual sensors optimized with automatic response. During the last month of each period, the following data concerning patient physical activity were retrieved from PM diagnostics (Activity Log): mean percentage of physical activity, mean intensity of activity. Quality of life (QoL) scores and 6-minute walk test (WT) were also recorded.

RESULTS

Single-sensor rate-responsive pacing resulted in symptomatic benefit equally with XL and MV sensors while no additional benefit was found using dual sensor. In a subgroup analysis, patients (17%) with marked chronotropic incompetence and with 0% atrial sensing received benefits from single sensor with an additional advantage from sensor (QoL: +21 +/- 14% P < 0.05; WT: +17 +/- 7% P < 0.02).

CONCLUSION

In most patients with rate-responsive devices, a single sensor is sufficient to achieve a satisfactory rate response. A dual sensor combination and optimization provides an additional benefit only in a selected population with an advanced atrial chronotropic disease.

Для цитирования: Padeletti Luigi, Pieragnoli Paolo, Di Biase Luigi, Colella Andrea, Landolina Maurizio, Moro Eugenio, Orazi Serafino, Vicentini Alfredo, Maglia Giampiero, Pensabene Orazio, Raciti Giovanni, Barold S Serge. Is a dual-sensor pacemaker appropriate in patients with sino-atrial disease? Results from the DUSISLOG study.. Pacing and clinical electrophysiology : PACE. 2006. DOI: 10.1111/j.1540-8159.2006.00301.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.