MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/01945998211006930

Медицинская библиотека · Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery · 2021

Adverse Events Associated With Corticosteroid-Eluting Sinus Stents: A MAUDE Database Analysis.

Narwani Vishal, Torabi Sina J, Kasle David A, Patel Rahul A, Lerner Michael Z, Manes R Peter

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Corticosteroid-eluting stents (CESs) are increasingly used after endoscopic sinus surgery to reduce the need for revision surgery, but their use is not without risks. The objective of this study is to describe adverse events related to CESs.

STUDY

DESIGN

Retrospective cross-sectional study.

SETTING

The US Food and Drug Administration's MAUDE database (2011-2020; Manufacturer and User Facility Device Experience).

METHODS

The MAUDE database was queried for reports of adverse events involving the use of CESs approved by the Food and Drug Administration, including Propel, Propel Mini, Propel Contour, and Sinuva (Intersect ENT).

RESULTS

There were 28 reported adverse events in total, with all events being related to the Propel family of stents and none related to Sinuva stents. Overall, 22 were categorized as patient-related adverse events and 6 as device-related events. The most common adverse event was related to postoperative infection, accounting for 39% (n = 11) of all complications. Four of these patients developed periorbital cellulitis, and 5 developed a fungal infection. The second-most common adverse event was migration of the stent, representing 21% of all complications (n = 6). Overall, 8 patients (29%) in our cohort required reintervention in the operating room, with subsequent removal of the CES.

CONCLUSION

The most commonly reported adverse events were postoperative infection, including multiple cases of fungal infection, followed by migration of the stent. An increased awareness of the complications associated with CESs can be used to better inform patients during the consenting process as well as surgeons in their surgical decision making.

Для цитирования: Narwani Vishal, Torabi Sina J, Kasle David A, Patel Rahul A, Lerner Michael Z, Manes R Peter. Adverse Events Associated With Corticosteroid-Eluting Sinus Stents: A MAUDE Database Analysis.. Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery. 2021. DOI: 10.1177/01945998211006930.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.