Effect of left ventricular function on long-term left ventricular pacing and sensing threshold.
Аннотация
BACKGROUND
The effect of left ventricular (LV) systolic function on the long-term left ventricular pacing and sensing threshold is unclear.
METHODS AND
RESULTS
We studied the effect of LV ejection fraction (LVEF) on the LV pacing and sensing threshold in 56 patients (mean age: 70.2 +/- 10.5 years) underwent permanent LV pacing using a self-retaining coronary sinus lead (Model 1055 K, St Jude Medical, USA). In 49 patients, the LV lead was implanted for conventional pacemaker indication (sick sinus syndrome = 14, heart block = 26 or slow atrial fibrillation = 9). The remaining 7 patients were implanted for congestive heart failure. The LV pacing and sensing threshold, and lead impedance were compared between patients with LVEF <40% (Group 1, n = 28) and LVEF >40% (Group 2, n = 28) during implant and at 3-month follow up. The LV pacing lead was successfully implanted in all patients without any lead dislodgement on follow-up. At implant, Group 1 patients had a significant lower R wave amplitude, but similar LV pacing threshold and lead impedance as compared to Group 2. However, at 3-month follow-up, Group 1 patients had a significantly higher LV pacing threshold compared to Group 2 patients. There were no significant differences in the sensing threshold and lead impedance between the two groups. Furthermore, there was also a significant interval increase in LV pacing threshold in Group 1 patients (0.94 +/- 0.12 V) after 3 months, but not in Group 2 patients (0.16 +/- 0.08 V, p < 0.01).
CONCLUSIONS
The results of this study suggest that the LV systolic function has a significant impact on the long-term LV pacing threshold. The long-term left ventricular pacing threshold in patients with left ventricular systolic dysfunction increased after implant and was higher than patients with normal left ventricular systolic function.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.