MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1046/j.1460-9592.2002.01274.x

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2002

Failure of atrial flutter detection by a pacemaker with a dedicated atrial flutter detection algorithm.

Israel Carsten W, Barold S Serge

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Detection of atrial flutter may be difficult for pacemakers with automatic mode switching algorithms. The Medtronic Kappa 700 device pacemakerfamily offers a dedicated Blanked Flutter Search algorithm specifically designed for the detection of atrial flutter. This report describes how spontaneous atrioventricular conduction, resulting in a sensed QRS complex at a critical time, inhibited the function of the Blanked Flutter Search algorithm to sense "concealed" or "blanked" atrialflutter. This observation underscores the need for further refinement of pacemaker algorithms for the detection of atrial tachyarrhythmias.

Для цитирования: Israel Carsten W, Barold S Serge. Failure of atrial flutter detection by a pacemaker with a dedicated atrial flutter detection algorithm.. Pacing and clinical electrophysiology : PACE. 2002. DOI: 10.1046/j.1460-9592.2002.01274.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.