[Delayed atrial excitation following bifocal pacemaker stimulation].

Alt E, Wirtzfeld A, Seidl K, Haller F, von Bibra H, Sauer E

Клинический случайАннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

A patient with drug-resistant ventricular tachycardia due to ischemic heart disease with severe left ventricular failure was successfully treated by the implantation of a DDD pacemaker system pacing at a rate of 90 beats/min (overdrive suppression). Additional therapy with high doses of beta-blockers was necessary. The ECG demonstrated a delay of 100 ms between atrial spike and p-wave. The hemodynamic effects of this ineffective atrial contraction were assessed by jugular venous puls tracing, phonocardiography, echocardiography, and radionuclide-ventriculography. The desired positive effects of physiological pacing could only be achieved by further prolongation of the A-V interval by these 100 ms. This observation shows that, with DDD pacemakers, AV intervals of varying length may be necessary with VAT or VDD and DVI modes in individual cases.

Для цитирования: Alt E, Wirtzfeld A, Seidl K, Haller F, von Bibra H, Sauer E. [Delayed atrial excitation following bifocal pacemaker stimulation].. Zeitschrift fur Kardiologie. 1983.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.