Whole-blood Point-of-Care Activated Partial Thromboplastin Time Ratio (APR) is not Accurate Enough to Monitor Heparin Therapy in Patients with Severe Respiratory Failure Secondary to SARS-Cov-2 Infection Supported with Veno-Venous Extracorporeal Membrane Oxygenation (VV-ECMO).

Aston Daniel, Besser Martin, Goddard Ben, Maggs Nicholas, Ahmed Hassan, Falter Florian

Наблюдательное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

Support with VV-ECMO requires anticoagulation with unfractionated heparin to prevent thrombotic complications. This must be monitored due to bleeding risk. A point-of-care (POC) method of testing aPTT and APR was evaluated for agreement with laboratory methods. In a prospective observational study, patients supported on VV-ECMO as a result of severe respiratory failure secondary to Covid-19 infection were given heparin as part of standard therapy. The aPTT was measured (i) at the bedside using the Hemochron Signature Elite device and (ii) at the hospital laboratory. Duplicate results were compared. Agreement between the POC and laboratory tests was poor, as assessed using the Bland-Altman method. The maximum difference between POC and laboratory methods was 133% and the minimum was 0%. Overall bias was 7.3% and limits of agreement were between -43.8% and 58.5%. Correlation increased when results were normalised to platelet count and creatinine. This POC test is insufficiently accurate for use as the primary method of heparin monitoring in patients requiring VV-ECMO for Covid-19. Platelets and renal function may influence the result of this whole blood POC test.

Для цитирования: Aston Daniel, Besser Martin, Goddard Ben, Maggs Nicholas, Ahmed Hassan, Falter Florian. Whole-blood Point-of-Care Activated Partial Thromboplastin Time Ratio (APR) is not Accurate Enough to Monitor Heparin Therapy in Patients with Severe Respiratory Failure Secondary to SARS-Cov-2 Infection Supported with Veno-Venous Extracorporeal Membrane Oxygenation (VV-ECMO).. Clinical and applied thrombosis/hemostasis : official journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis. 2022. DOI: 10.1177/10760296221148162.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.