Analytical Validation of an Automated Point-of-Care Immunoassay for the Measurement of Canine Pancreatic Lipase Immunoreactivity Concentration (Vcheck cPL 2.0).

Mendoza-White Isabel, Steiner Joerg M, Cridge Harry

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Pancreatic lipase assays are commonly utilized in the diagnostic approach to suspected pancreatitis in dogs. While a previously available point-of-care assay for the quantification of canine pancreatic lipase immunoreactivity (Vcheck cPL 1.0) performed sub-optimally, the updated version (Vcheck cPL 2.0) lacks independent validation.

OBJECTIVE

To evaluate the analytic validity of the Vcheck cPL 2.0 assay.

METHODS

Linearity was assessed via dilutional parallelism. Intra- and inter-assay variability were assessed by calculating coefficients of variation across replicates on the same day and across consecutive days, respectively. Interference studies were performed to assess the effects of bilirubin, hemoglobin, and lipid presence in test samples on assay performance.

RESULTS

Linearity appeared within target parameters at pancreatic lipase concentrations up to ~800 md/dL (mean O/E ratio: 0.968 ± 0.104) but was unacceptable at pancreatic lipase concentrations > 2000 mg/dL (mean O/E ratio: 1.432 ± 0.178). Intra-assay precision (median coefficient of variation 7.2%) was acceptable; however, inter-assay precision was sub-optimal (coefficient of variation > 10%) in 50% of samples analyzed. Variable precision may result in different diagnostic interpretations. Icterus and lipemia did not impact assay results. However, hemolysis, when severe (375-500 mg/dL), may impact test results (p < 0.005).

CONCLUSIONS

The Vcheck cPL 2.0 shows improved precision compared to previous versions. Linearity was unacceptable at higher pancreatic lipase concentrations and hemolysis may impact Vcheck cPL 2.0 assay results.

Для цитирования: Mendoza-White Isabel, Steiner Joerg M, Cridge Harry. Analytical Validation of an Automated Point-of-Care Immunoassay for the Measurement of Canine Pancreatic Lipase Immunoreactivity Concentration (Vcheck cPL 2.0).. Veterinary clinical pathology. 2026. DOI: 10.1111/vcp.70106.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.