Quantitative Comparison Of Microvascular Metrics On Three Optical Coherence Tomography Angiography Devices In Chorioretinal Disease.

Lu Yifan, Wang Jay C, Zeng Rebecca, Katz Raviv, Vavvas Demetrios G, Miller Joan W, Miller John B

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Optical coherence tomography angiography (OCT-A) has emerged as a novel tool for the non-invasive imaging and evaluation of the retinal microvasculature. There is little existing literature that compares OCT-A microvasculature metrics across different OCT-A devices in chorioretinal diseases. Herein, we examined these metrics on three available OCT-A platforms.

PATIENTS AND

METHODS

All subjects were scanned on each of three OCT-A devices: Optovue Avanti Angiovue, Topcon DRI-OCT Triton Swept-Source OCT, and Zeiss Cirrus 5000-HD-OCT Angioplex. Two investigators independently measured foveal avascular zone (FAZ) area. Superficial capillary plexus (SCP) and deep capillary plexus (DCP) vessel densities (VD) were calculated from binarized images with ImageJ software. Image quality across devices was qualitatively compared. Interclass correlation coefficient (ICC), Bland-Altman analysis, repeated measures ANOVA, and post-hoc tests were performed for statistical analysis.

RESULTS

Thirteen eyes of seven patients with chorioretinal diagnoses were reviewed. ICC for FAZ measurement was 0.95. There was no significant difference in FAZ area across the three devices (p= 0.792). There was a significant difference in the SCP VD between the OCT-A devices (Triton 0.344 ± 0.013, Angiovue 0.323 ± 0.013, Angioplex 0.367 ± 0.014, p < 0.05). Significantly greater DCP VD was observed with Angioplex (0.385 ± 0.010) in comparison to both Triton (0.331 ± 0.009) and Angiovue (0.341 ± 0.020). A comparison of image quality revealed that Angiovue gives the highest quality, followed by Angioplex and Triton.

CONCLUSION

Core macular microvasculature metrics are now readily accessible on a variety of available OCT-A devices. While the FAZ can be reliably measured across all three devices in this study, there were significant differences for the vessel density in both the SCP and DCP. As a result, clinicians should be careful when comparing microvasculature metrics across different devices when using patient data in multicenter research investigations and clinical trials.

Для цитирования: Lu Yifan, Wang Jay C, Zeng Rebecca, Katz Raviv, Vavvas Demetrios G, Miller Joan W, Miller John B. Quantitative Comparison Of Microvascular Metrics On Three Optical Coherence Tomography Angiography Devices In Chorioretinal Disease.. Clinical ophthalmology (Auckland, N.Z.). 2019. DOI: 10.2147/opth.s215322.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.