Comparison of a viscoelastic point-of-care coagulation monitor with thromboelastography in sick dogs with hemostatic abnormalities.

Buriko Yekaterina, Chalifoux Nolan V, Clarkin-Breslin Rachel, Silverstein Deborah C

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Viscoelastic coagulation monitor (VCM-Vet) is a point-of-care device that has been used to characterize hemostatic abnormalities in sick pets but has not been validated in veterinary patients.

OBJECTIVES

We aimed to compare VCM-Vet and thromboelastography (TEG) in sick dogs with suspected disorders of hemostasis.

METHODS

Duplicate VCM-Vet tests using untreated native blood performed concurrently on two VCM-Vet machines, and simultaneous TEG tests were performed (one citrated native (CN), and one activated with tissue factor (TF) at a 1:3600 dilution). Each VCM-Vet result was compared with both TF-activated and CN TEG.

RESULTS

Fifty-three dogs were enrolled. Eleven cases displayed apparent hyperfibrinolysis. Spearman correlation coefficients for individual VCM-Vet devices and CN and TF TEG were obtained between R and CT values and ranged from 0.21 to 0.27, CFT and K (r = 0.60-0.67), angles (r = 0.51-0.62), and MCF and MA (r = 0.85-0.87). Comparison of the two VCM-Vet devices displayed positive correlations for all clot formation parameters with Lin's concordance correlation coefficients of 0.75-0.95. Variable lysis parameter agreement existed between the VCM-Vet devices and VCM-Vet and TEG. When samples were classified as hypercoagulable or coagulopathic, VCM-Vet had a low positive predictive value (17-33%) for the detection of hypercoagulable states and a moderate negative predictive value (64-74%) for the detection of coagulopathy as defined by TEG.

CONCLUSIONS

VCM-Vet and TEG had variable correlations in clot formation values and a strong correlation for final clot strength. More information is needed to make conclusions about the lysis parameters. Artifact in the fibrinolysis portion of the test can confound the interpretation of VCM-Vet results.

Для цитирования: Buriko Yekaterina, Chalifoux Nolan V, Clarkin-Breslin Rachel, Silverstein Deborah C. Comparison of a viscoelastic point-of-care coagulation monitor with thromboelastography in sick dogs with hemostatic abnormalities.. Veterinary clinical pathology. 2023. DOI: 10.1111/vcp.13198.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.