Novel Specturi urine collector device demonstrating high measurement accuracy and proven clinical usability.
Аннотация
OBJECTIVES
Urine collection from neonates and infants remains challenging. Non-invasive urine collection methods (UCMs) are preferred, but they are often associated with drawbacks, such as feces contamination, discomfort, and/or inaccurate urinary analysis. This study aimed to demonstrate the measurement accuracy and clinical usability of the Specturi urine collector device.
METHODS
25 different urines were incubated in the Specturi urine collector device and three non-invasive UCMs, namely gauzes, urine collection bag, and PeeSpot, for 3 h at 37 °C. Albumin, calcium, chloride, creatinine, glucose, magnesium, phosphate, potassium, protein, sodium, and urea concentrations as well as osmolality and pH were measured in the extracted urines and compared to a control. The measurements were also compared between the different UCMs. In addition, a clinical pilot study was performed with the Specturi urine collector device to assess the usability and safety as indicated by nurses and parents.
RESULTS
The novel Specturi urine collector device offered accurate measurements of the various urinary analytes. Its measurement accuracy was comparable to the urine collection bag and PeeSpot and more accurate than gauzes. Nurses indicated that the Specturi urine collector device was easy to use, noting the ease of application and removal. Both nurses and parents indicated that the participants experienced no discomfort while wearing the Specturi urine collector device and no significant skin reactions were observed, unlike for other UCMs.
CONCLUSIONS
The Specturi urine collector device is a novel non-invasive UCM, offering accurate urinary analysis, ensuring patient safety and comfort, and providing easy usage by nurses.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.