A New Anti-Parietal Cell Antibody Titer Measurement Kit for Diagnosis of Autoimmune Gastritis.
Аннотация
UNLABELLED: <p>
Introduction
Recent findings revealed that autoimmune gastritis (AIG) is not rare in Japan. Therefore, the accurate diagnosis of AIG is essential in upper gastrointestinal practice. Diagnostic criteria for AIG were established in 2023; however, the conventional fluorescent antibody (FA) method for anti-parietal cell antibody (APCA) titer measurements has low accuracy and is clinically problematic.
METHODS
We developed a latex agglutination (LA) method using a new human-derived antigen. Samples subjected to measurements were sera from AIG cases (127 cases, 49 males, average age 65.9 years) and control cases (129 cases, 81 males, average age 66.1 years) provided by the main facility and three affiliated facilities in Japan. APCA in whole serum was measured using the FA and LA methods.
RESULTS
The diagnostic ability of AIG using the FA method was as follows: sensitivity of 99.2%, specificity of 41.1%, and an overall agreement rate of 69.9%, with low specificity previously being reported. On the other hand, the new LA method had good diagnostic performance with a sensitivity of 93.7%, a specificity of 95.4%, and an overall agreement rate of 94.5%.
CONCLUSION
We developed a new APCA measurement system that will contribute to the accurate diagnosis of AIG. The use of this measurement system in clinical practice will facilitate the diagnosis and treatment of AIG. </p>.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.