DOI 10.1002/jmri.29383

Медицинская библиотека · Journal of magnetic resonance imaging : JMRI · 2024

Second-Order Motion-Compensated Echo-Planar Cardiac Diffusion-Weighted MRI: Usefulness of Compressed Sensitivity Encoding.

Chen Rui, Luo Ruohong, Xu Yongzhou, Ou Jiehao, Li Xiaodan, Yang Yuelong, Cao Liqi, Wu Zhigang, Luo Wei, Liu Hui

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Cardiac diffusion-weighted imaging (DWI) using second-order motion-compensated spin echo (M2C) can provide noninvasive in-vivo microstructural assessment, but limited by relatively low signal-to-noise ratio (SNR). Echo-planar imaging (EPI) with compressed sensitivity encoding (EPICS) could address these issues.

PURPOSE: To combine M2C DWI and EPCIS (M2C EPICS DWI), and compare image quality for M2C DWI.

STUDY TYPE: Prospective.

POPULATION: Ten ex-vivo hearts, 10 healthy volunteers (females, 5 [50%]; mean ± SD of age, 25 ± 4 years), and 12 patients with diseased hearts (female, 1 [8.3%]; mean ± SD of age, 44 ± 16 years; including coronary artery heart disease, congenital heart disease, dilated cardiomyopathy, amyloidosis, and myocarditis).

FIELD STRENGTH/SEQUENCE: 3-T, M2C EPICS DWI, and M2C DWI.

ASSESSMENT: The apparent SNR (aSNR) and the rating scores were used to evaluate and compared image quality of all three groups. The aSNR was calculated using aSNR = Mean intensity myocardium / Standard deviation myocardium , and the myocardium was segmented manually. Three observers independently rated subjective image quality using a 5-point Likert scale.

STATISTICAL TESTS: Bland-Altman analysis and paired t-tests. The threshold for statistical significance was set at P < 0.05.

RESULTS

In healthy volunteers, the aSNR with a b-value of 450 s/mm2 acquired by M2C EPICS DWI was significantly higher than M2C DWI at in-plane resolutions of 3.0 × 3.0, 2.5 × 2.5, and 2.0 × 2.0 mm2. In patients with diseased hearts, the aSNR ofM2C EPICS DWI was also significantly higher than that for M2C DWI (bias of M2C EPICS-M2C = 1.999, 95% limits of agreement, 0.362 to 3.636; mean ± SD, 7.80 ± 1.37 vs. 5.80 ± 0.81). The ADC values of M2C EPICS was significantly higher than M2C DWI in in-vivo hearts. Over 80% of the images with rating scores for M2C EPICS DWI were higher than M2C DWI in in-vivo hearts.

DATA

CONCLUSION

Cardiac imaging by M2C EPICS DWI may demonstrate better overall image quality and higher aSNR than M2C DWI.

EVIDENCE LEVEL: 2 TECHNICAL EFFICACY: Stage 1.

Для цитирования: Chen Rui, Luo Ruohong, Xu Yongzhou, Ou Jiehao, Li Xiaodan, Yang Yuelong, Cao Liqi, Wu Zhigang, Luo Wei, Liu Hui. Second-Order Motion-Compensated Echo-Planar Cardiac Diffusion-Weighted MRI: Usefulness of Compressed Sensitivity Encoding.. Journal of magnetic resonance imaging : JMRI. 2024. DOI: 10.1002/jmri.29383.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.