Correlation Between Corneal Topographic, Densitometry, and Biomechanical Parameters in Keratoconus Eyes.

Shen Yang, Han Tian, Jhanji Vishal, Shang Jianmin, Zhao Jing, Li Meiyan, Zhou Xingtao

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the correlation between corneal densitometry, corneal topographic parameters, and corneal biomechanical properties in keratoconus.

METHODS

A total of 76 eyes of 76 keratoconus patients were enrolled in this cross-sectional study. Corneal densitometry and topography were measured using Pentacam HR. Corneal biomechanical properties were measured using CorVis ST.

RESULTS

The corneal densitometry values of the anterior 0 to 2 and 2 to 6 mm layers significantly correlated with the maximum keratometry values (R = 0.373, P = 0.001 and R = 0.276, P = 0.016, respectively), thinnest corneal thickness values (R = -0.331, P = 0.003 and R = -0.234, P = 0.042, respectively), anterior corneal elevation (R = 0.392, P < 0.001 and R = 0.323, P = 0.004, respectively), and posterior corneal elevation (R = 0.450, P < 0.001 and R = 0.367, P = 0.001, respectively). The stiffness parameter-applanation time 1 (SP-A1) significantly correlated with the corneal densitometry values for the anterior 0 to 2 mm (R = -0.397, P < 0.001), anterior 2 to 6 mm (R = -0.331, P = 0.004), central 0 to 2 mm (R = -0.306, P = 0.007), central 2 to 6 mm (R = -0.228, P = 0.048), posterior 2 to 6 mm (R = -0.243, P = 0.035), total 0 to 2 mm (R = -0.291, P = 0.011), and total 2 to 6 mm (R = -0.295, P = 0.010) layers.

CONCLUSIONS

The corneal densitometry values correlated with the severity of keratoconus and the SP-A1 values.

TRANSLATIONAL RELEVANCE: Corneal densitometry values may serve as markers to predict the severity of keratoconus.

Для цитирования: Shen Yang, Han Tian, Jhanji Vishal, Shang Jianmin, Zhao Jing, Li Meiyan, Zhou Xingtao. Correlation Between Corneal Topographic, Densitometry, and Biomechanical Parameters in Keratoconus Eyes.. Translational vision science & technology. 2019. DOI: 10.1167/tvst.8.3.12.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.