DOI 10.1016/j.jcrs.2019.02.005

Медицинская библиотека · Journal of cataract and refractive surgery · 2019

Corneal higher-order aberrations measurements: Comparison between Scheimpflug and dual Scheimpflug-Placido technology in keratoconic eyes.

Piccinini Andre L, Golan Oren, Torres-Netto Emilio A, Hafezi Farhad, Randleman J Bradley

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the difference and agreement of corneal higher-order aberrations (HOAs) in keratoconic eyes using Scheimpflug and dual Scheimpflug-Placido imaging systems.

SETTING

Emory University, Atlanta, Georgia, and the University of Southern California, Los Angeles, California, USA.

DESIGN

Retrospective between-devices reliability and agreement study.

METHODS

Patients diagnosed with keratoconus were evaluated sequentially by Scheimpflug and dual Scheimpflug-Placido devices. Differences, correlations, and agreement between values for total root mean square (RMS), trefoil, coma, and spherical aberration were analyzed, and Bland-Altman plots were generated.

RESULTS

Fifty eyes from 50 patients (31 men, 19 women) were evaluated. Trefoil at 30 degrees, spherical aberration, and total RMS were significantly different between groups (P < .05), whereas trefoil at 0 degrees and total coma values were not statistically different. There was a weak positive correlation between devices for trefoil at 0 degrees (r = 0.228), and a moderate positive correlation for trefoil at 30 degrees (r = 0.473), horizontal coma (r = 0.430), and for total corneal RMS (r = 0.637). Vertical coma (r = 0.816) and spherical aberration (r = 0.874) showed a strong positive correlation. The 95% limits of agreement (LoA) for absolute values were 1.963 μm for trefoil at 30 degrees, 2.449 μm for trefoil at 0 degrees, 3.530 μm for horizontal coma, 2.145 μm for vertical coma, 1.242 μm for spherical aberration, and 10.527 μm for RMS.

CONCLUSION

Significant differences were found between measurements of corneal HOAs generated by Scheimpflug and dual Scheimpflug-Placido devices in patients with keratoconus, with generally limited correlations and wide LoA. HOAs measurements from these devices should not be considered equivalent.

Для цитирования: Piccinini Andre L, Golan Oren, Torres-Netto Emilio A, Hafezi Farhad, Randleman J Bradley. Corneal higher-order aberrations measurements: Comparison between Scheimpflug and dual Scheimpflug-Placido technology in keratoconic eyes.. Journal of cataract and refractive surgery. 2019. DOI: 10.1016/j.jcrs.2019.02.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.