MedElevator Войти в исследования
← Поиск публикаций

DOI 10.24875/cirue.m18000007

Медицинская библиотека · Cirugia y cirujanos · 2019

Reproducibility of the measurement of central corneal thickness in healthy subjects obtained with the optical low coherence reflectometry pachymeter and comparison with the ultrasonic pachymetry.

Garza-Leon Manuel, Plancarte-Lozano Eduardo, Valle-Penella Agustín Del, Guzmán-Martínez María de Lourdes, Villarreal-González Andrés

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Corneal pachymetry is widely used for refractive surgery and follow up in keratoconus, accurate measurement is essential for a safe surgery.

OBJECTIVE

To assess intraobserver reliability of central corneal thickness (CCT) measurements using optical low-coherence reflectometry (OLCR) technology and its agreement with ultrasonic pachymeter (US).

METHOD: Randomized and prospective comparative evaluation of diagnostic technology. One randomly healthy eye of subjects was scanned three times with both devices. Intraobserver within-subject standard deviation (Sw), coefficient of variation (CVw) and intraclass correlation coefficient (ICC) were obtained for reliability analysis; for study agreement, data were analyzed using the paired-sample t test and the Bland-Altman LoA method. The mean of three scans of each equipment was used to assess the LoA.

RESULTS

The study enrolled 30 eyes of 30 subjects with average age of 28.70 ± 8.06 years. For repeatability, the Sw were 3.41 and 5.96 µ, the intraobserver CVw was 2 and 4% and ICC 0.991 and 0.988, for OLCR and US respectively. The mean CCT difference between OLCR and US was 8.90 ± 9.03 µ (95% confidence interval: 5.52-2.27 µ), and the LoA was 35.40 µ.

CONCLUSIONS

OLCR technology provided reliable intraobserver CCT measurements. Both pachymetry measurements may be used interchangeably with minimum calibration adjustment.

Для цитирования: Garza-Leon Manuel, Plancarte-Lozano Eduardo, Valle-Penella Agustín Del, Guzmán-Martínez María de Lourdes, Villarreal-González Andrés. Reproducibility of the measurement of central corneal thickness in healthy subjects obtained with the optical low coherence reflectometry pachymeter and comparison with the ultrasonic pachymetry.. Cirugia y cirujanos. 2019. DOI: 10.24875/cirue.m18000007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.