Accuracy in tumor size measurements: Comparison of digital mammography, digital breast tomosynthesis and synthetic mammography.

Şendur Halit Nahit, Cerit Mahi Nur, Gültekin Serap, Cindil Emetullah, Avdan Aslan Aydan, Erdal Zeynep Sezgi, Gültekin Işıl İmge, Teke Furkan

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The aim of this study was to assess the accuracy of digital mammography (DM), digital breast tomosynthesis (DBT) and synthetic mammography (SM) in tumor size measurements compared with histological tumor sizes.

MATERIALS AND

METHODS

71 breast cancer patients who underwent DM and DBT acquisitions simultaneously were included in this study. One radiologist with 8 years of experience in breast imaging measured tumor sizes independently in three separate sessions which include DM, DBT and SM images, respectively. The correlations between the measured tumor sizes on each imaging technique and histological sizes were analyzed using Spearman correlation test. The patients were categorized into two subgroups according to assigned breast density categories (dense and non-dense), and histological tumor sizes (≤2 cm and > 2 cm). To assess the agreement levels between the measured tumor sizes and histological sizes Bland-Altman analyses were performed for each imaging technique.

RESULTS

The mean of histological size of tumors was 23.85 ± 16.57 mm (median: 20). The means of measured tumor sizes were 21.21 ± 13.59 mm (median: 19), 21.52 ± 13.42 mm (median: 19) and 18.97 ± 11.21 mm (median: 17) in DM, DBT and SM, respectively. The Spearman correlation values with histologic sizes were 0.814 (P < 0.001), 0.887 (P < 0.001), and 0.852 (P < 0.001) for DM, DBT and SM, respectively. In subgroup analyses, the correlation values showed decrement for tumors >2 cm in size compared to tumors ≤2 cm in size.

CONCLUSION

DBT provides the most accurate tumor size measurements among mammographic imaging techniques and if mammography will be used in tumor size measurements, DBT should be preferred.

Для цитирования: Şendur Halit Nahit, Cerit Mahi Nur, Gültekin Serap, Cindil Emetullah, Avdan Aslan Aydan, Erdal Zeynep Sezgi, Gültekin Işıl İmge, Teke Furkan. Accuracy in tumor size measurements: Comparison of digital mammography, digital breast tomosynthesis and synthetic mammography.. Clinical imaging. 2020. DOI: 10.1016/j.clinimag.2020.07.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.