MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/17576180.2024.2403241

Медицинская библиотека · Bioanalysis · 2024

Development and validation of an ultra-performance liquid chromatography with tandem mass spectrometry method for determination of soluble repulsive guidance molecule A in human serum and cerebrospinal fluid.

Schreier Patrick, Huang Lili, Fung Emma, Mollon Jennifer, Sielaff Bernhard, Lake Marc R, Schulz Michael, Awwad Khader

СтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

Aim: Repulsive guidance molecule A (RGMa) is upregulated in neurodegenerative diseases. To assess RGMa levels in human serum and cerebrospinal fluid (CSF), a quantification method was developed and validated according to ICH M10 guideline.Methods & results: Sample preparation consisted of immunoprecipitation (IP, only for serum), digestion and purification followed by MS.

Conclusion

An UPLC-MS/MS method was established and used to assess normal range of soluble RGMa levels in serum and CSF of healthy controls, and patients with mild cognitive impairment or Alzheimer's disease. The normal range was between 13.0-44.8 ng/ml (CSF) and 9.9-20.9 ng/ml (serum) in healthy controls. In the CSF of patients with mild cognitive impairment and Alzheimer's disease, total soluble RGMa was twofold lower while unchanged in serum.

Для цитирования: Schreier Patrick, Huang Lili, Fung Emma, Mollon Jennifer, Sielaff Bernhard, Lake Marc R, Schulz Michael, Awwad Khader. Development and validation of an ultra-performance liquid chromatography with tandem mass spectrometry method for determination of soluble repulsive guidance molecule A in human serum and cerebrospinal fluid.. Bioanalysis. 2024. DOI: 10.1080/17576180.2024.2403241.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.