Comparison of ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) and enzyme-multiplied immunoassay technique (EMIT) for quantification of voriconazole plasma concentration from Chinese patients.

Yu Mingjie, Yang Jun, Xiong Lirong, Zhan Shipeng, Cheng Lin, Chen Yongchuan, Liu Fang

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Voriconazole (VRZ) is the recommended standard treatment for life-threatening invasive aspergillosis. The plasma concentration of VRZ should be determined to optimise treatment results and reduce side effects. This study aimed to compare the correlation and concordance of ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) and enzyme-multiplied immunoassay technique (EMIT) to determine VRZ plasma concentration in clinical practice.

METHODS

An isotopically labelled internal standard UPLC-MS/MS method was established, validated, and subsequently applied to determine VRZ concentration. The UPLC-MS/MS method was also compared with a commercial EMIT method regarding results correlation and concordance.

RESULTS

The calibration curve of UPLC-MS/MS was linear from 0.1 to 10 mg/L, the inter- and intra-day relative standard deviations (RSDs), and the stability of quality control samples were less than 15 %, satisfying the Bioanalytical Method Validation Guidelines. A total of 122 plasma samples were collected and analyzed using both methods. UPLC-MS/MS and EMIT showed a high correlation (r = 0.9534), and Bland-Altman analysis indicated a mean absolute bias of 1.035 mg/L and an average bias of 27.56 % between UPLC-MS/MS and EMIT. The paired Wilcoxon test and Bland-Altman analysis revealed poor consistency between the two methods. Furthermore, we compared the effects of different methods in clinical applications. Two threshold values for treatment efficacy (1.0 mg/L) and safety (5.5 mg/L) were established, and considerable discordance was observed between the original EMIT and UPLC-MS/MS results at both thresholds (p < 0.05). Nevertheless, the adjusted EMIT results were not inconsistent with the UPLC-MS/MS results regarding the efficacy (p = 0.125) and safety (p = 1.0) thresholds.

CONCLUSIONS

The isotopically labelled internal standard UPLC-MS/MS method is established and well applied in the clinical setting. A strong correlation but discordance was found between UPLC-MS/MS and EMIT, indicating that switching from UPLC-MS/MS to EMIT was unsuitable. However, the adjusted EMIT results may serve as a reliable surrogate when UPLC-MS/MS results cannot be obtained when necessary.

Для цитирования: Yu Mingjie, Yang Jun, Xiong Lirong, Zhan Shipeng, Cheng Lin, Chen Yongchuan, Liu Fang. Comparison of ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) and enzyme-multiplied immunoassay technique (EMIT) for quantification of voriconazole plasma concentration from Chinese patients.. Heliyon. 2023. DOI: 10.1016/j.heliyon.2023.e22015.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.