[A preliminary study on the tear matrix metalloproteinase 9 point-of-care assay using a domestic kit].
Аннотация
Objective
To compare the point-of-care assays for tear matrix metalloproteinase 9 (MMP-9) using domestic and InflammaDry kits, and to evaluate the feasibility of diagnosing dry eye with the domestic kit.
Methods
It was a cross-sectional study. Thirty dry eye patients and 30 age-and sex-matched normal volunteers were continuously enrolled in this cross-sectional study from June 2022 to July 2022. Both domestic and InflammaDry kits were used to detect the tear MMP-9 levels. The positive rates were recorded for qualitative analysis, and the gray ratios of bands (the gray value of detection bands to that of control bands) were collected for quantitative analysis. The correlations of MMP-9 levels with age, ocular surface disease index, fluorescence tear break-up time, tear meniscus height, Schirmer's Ⅰ test score, corneal fluorescein staining score, and meibomian gland dropout were analyzed. The Mann-Whitney U test, paired Chi-square test, Kappa test, and Spearman's correlation coefficient were used for statistical analysis.
Results
There were 14 males and 16 females (30 eyes) in the control group, and their age was (39.37±19.55) years. In the dry eye group, 11 males and 19 females (30 eyes), aged (46.87±17.85) years, had moderate to severe dry eye. The positive rates of MMP-9 in tear fluid were significantly different between dry eye patients (InflammaDry: 86.67%; domestic kit: 70.00%) and controls (InflammaDry: 16.67%, P<0.001; domestic kit: 6.67%, P<0.001). Although the sensitivity of the domestic kit was lower than that of the InflammaDry kit (70.0% vs. 86.7%, P=0.001), the specificity was higher (93.3% vs. 83.3%, P=0.001). In dry eye patients, the positive coincidence rate was 80.7% (21/26), the negative coincidence rate was 100% (4/4), and the total coincidence rate was 83.3% (25/30), with no significant difference between the two kits (McNemar test: χ2=3.20, P>0.05), and the results of both kits were consistent (Kappa=0.53, P=0.001). The Spearman's correlation coefficient showed the gray ratios using both kits were positively correlated with the corneal fluorescein staining score (InflammaDry: ρ=0.48, P<0.05; domestic kit: ρ=0.52, P=0.003).
Conclusion
The performances of the domestic and InflammaDry kits are consistent in the point-of-care assay for tear MMP-9, and the domestic kit has lower sensitivity but higher specificity.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.