Diagnostic accuracy of non-invasive tear film break-up time assessed by the simple manual interferometric device.

Vidas Pauk Sania, Petriček Igor, Tomić Martina, Bulum Tomislav, Jandroković Sonja, Pauk Gulić Maja, Kalauz Miro, Lešin Gaćina Dina

СтатьяПеревод пока не подготовлен

Аннотация

AIM: To determine the diagnostic accuracy of non-invasive tear film break-up time (NIBUT) measured by the handheld lipid layer examination instrument.

METHODS

108 patients were enrolled in this cross-sectional study and divided into two groups: patients with dry eye (n = 57) categorized by the presence of dry eye symptoms obtained by Schein Questionnaire and minimally-one objective dry eye sign (tear film break-up time <10 s or corneal, conjunctival and lid margin fluorescein staining), and healthy subjects (n = 51).

RESULTS

Dry eye subjects had significantly shorter NIBUT than healthy subjects (6 s vs 20 s, p < 0.001). Logistic regression analysis showed that shorter NIBUT values were excellent indicators of dry eye disease (p < 0.001), with consistency and no significant difference between measurements, even after standardizing the results for age and sex. NIBUT cut-off point to distinguish dry eye from healthy subjects was 12 s (sensitivity 90.2 %, specificity 88.5 %, PPV 92.5 %, NPV 85.2 %, LR +7.82, LR- 0.11, DOR 70.92, DE 89.6 %). Good, but lower accuracy was observed at cut-off value of 10 s (sensitivity 87.8 %, specificity 88.5 %, PPV 92.3 %, NPV 82.1 %, LR+ 7.61, LR- 0.14, DOR 55.2, DE 88.1 %). The area under the ROC curve (AUC) of 0.944 classified NIBUT as a diagnostic test with very high accuracy.

CONCLUSION

This study showed a high diagnostic accuracy of NIBUT measured by the handheld lipid layer examination instrument. This simple, reliable, objective and available instrument might regularly take place in routine, standard dry eye diagnostic and can be used by almost every eye specialist.

Для цитирования: Vidas Pauk Sania, Petriček Igor, Tomić Martina, Bulum Tomislav, Jandroković Sonja, Pauk Gulić Maja, Kalauz Miro, Lešin Gaćina Dina. Diagnostic accuracy of non-invasive tear film break-up time assessed by the simple manual interferometric device.. Contact lens & anterior eye : the journal of the British Contact Lens Association. 2022. DOI: 10.1016/j.clae.2022.101776.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.