Impact of Median Sternotomy on Safety and Efficacy of the Subcutaneous Implantable Cardioverter Defibrillator.

Sugrue Alan, Ibrahim Rand, Lu Marvin, Bhatia Neal K, Alkukhun Laith, Adewumi Joseph, Schaller Robert D, Marchlinski Francis E, D'Souza Benjamin, Nazer Babak, Tzou Wendy, Merchant Faisal M, Frankel David S

Многоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Subcutaneous implantable cardioverter defibrillators (S-ICDs) are an attractive alternative to transvenous ICDs among those not requiring pacing. However, the risks of damage to the S-ICD electrode during sternotomy and adverse interactions with sternal wires remain unclear. We sought to determine the rates of damage to the S-ICD lead during sternotomy, inappropriate shocks from electrical noise due to interaction with sternal wires, and failure to terminate spontaneous or induced ventricular arrhythmias.

METHODS

Retrospective, multicenter study of patients undergoing sternotomy before or after S-ICD implantation. Clinical, procedural, and device-related data were collected by each center and analyzed by the coordinating center. These data were compared with a historical control cohort of nonsternotomy patients.

RESULTS

Of 196 identified patients (52±16 years, 47 women), 166 underwent S-ICD implantation after sternotomy and 30 sternotomy after S-ICD. There was no damage to any lead among those who underwent sternotomy after S-ICD. Defibrillation threshold testing was performed in 63% at implant, with 91% first shock success. During a median follow-up of 29 months (range, 1-188), S-ICD first shocks successfully terminated spontaneous ventricular arrhythmias in 31 of 32 patients (97%). Inappropriate shocks occurred in 22 patients, most commonly related to T wave oversensing (n=14). Compared with the nonsternotomy controls, there were no differences in rates of first shock success for induced or spontaneous arrhythmias or rate of inappropriate shocks.

CONCLUSIONS

Sternotomy before or after S-ICD does not confer additional risk relative to a historical control group without sternotomy.

Для цитирования: Sugrue Alan, Ibrahim Rand, Lu Marvin, Bhatia Neal K, Alkukhun Laith, Adewumi Joseph, Schaller Robert D, Marchlinski Francis E, D'Souza Benjamin, Nazer Babak, Tzou Wendy, Merchant Faisal M, Frankel David S. Impact of Median Sternotomy on Safety and Efficacy of the Subcutaneous Implantable Cardioverter Defibrillator.. Circulation. Arrhythmia and electrophysiology. 2023. DOI: 10.1161/circep.123.011867.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.