DOI 10.23736/s0021-9509.24.13039-x
Safety and efficacy of the automated suture system in valvular heart surgery: a multicenter, prospective registry.
Аннотация
BACKGROUND
In the last years, the Cor-Knot® device has been increasingly used in heart valve surgery. Our aim was to investigate the incidence of valvular complications in patients who underwent valvular surgery using the Cor-Knot® device in multicentric cohorts at one-year follow-up.
METHODS
Three hundred and sixty-eight patient underwent heart valve repair or replacement surgery using automated titanium suture fasteners in four cardiothoracic surgery departments between September 2018 and January 2020.
RESULTS
The mean age was 66.3±10.2 years. The mean Euroscore II was 3.14±5.65. Procedures were performed by right anterior mini-thoracotomy for 264 patients (71.7%) and by conventional sternotomy for 76 (20.7%). Isolated single valve surgery was most common (285 patients, 77.4%), 31 patients (8.4%) underwent isolated double valve surgery and 6 patients isolated triple valve surgery (1.6%). An associated procedure was performed in 46 patients (12.5%). Sixteen patients (4.3%) required permanent pacemaker implantation in the postoperative period. Eighteen patients died postoperatively (4.9%). Two patients had paravalvular leak ≥2 (0.5%). Mean follow-up was 14.7±7.1 months. Eight patients died during the follow-up (2.2%). Four patients had infectious endocarditis. The rate of valvular leak ≥2 was 1.5% and four patients underwent valve-related reoperation (1.2%), no reoperation was related to a valve lesion due to Cor-Knot®. There were three pacemaker implantation (0.9%) and 12 patients had NYHA>2 (3.7%). No case of metallic embolization, prosthesis thrombosis or leaflet perforation was reported during follow-up.
CONCLUSIONS
The use of the Cor-Knot® automated knotting system in valve surgery is not associated with an increase in the rate of paravalvular leakage, permanent pacemaker implantation or mortality postoperatively or during follow-up.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.