Thermal damage of the specimen during breast biopsy with the use of the Breast Lesion Excision System: does it affect diagnosis?
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Al-Harethee Wasim Ali, Kalles Vasileios, Papapanagiotou Ioannis, Matiatou Maria, Georgiou Georgia, Nonni Afrodite, Koulocheri Dimitra, Liakou Paraskevi, Theodoropoulos George, Zografos George C
Аннотация
BACKGROUND: Stereotactic vacuum assisted breast biopsy (VABB) procedures remain of key importance in the diagnostic evaluation of suspicious non-palpable mammographic lesions. The Breast Lesion Excision System(®) (BLES) is an image-guided percutaneous biopsy method that utilizes radiofrequency (RF) in order to retrieve an intact-tissue biopsy specimen. The purpose of this study is to determine the effects of RF application on the tissue specimen and the possible interference with the histopathologic results.
METHODS: In this study we included 265 patients with suspicious non-palpable mammographic lesions who underwent 273 stereotactic VABB procedures using the BLES. All biopsies were performed by the same surgical-radiology team under local anesthesia. The specimen thermal damage was classified according to the pathology report in 3 categories: Severe (extensive thermal damage or inability to diagnose), medium (ability to diagnose but either circumferential damage >1.5 mm or diffuse areas of thermal damage) and mild (circumferential thermal damage <1.5 mm but >0.5 mm).
RESULTS: Radiofrequency-associated thermal damage of the specimen was observed in 14 cases (5.13 %), and was classified as severe in 5, medium in 5, and mild in 4 specimens. Within the group of RF damaged specimens, we found a significant (p < 0.05) positive correlation between fat cell content and classification of thermal damage.
CONCLUSIONS: Although thermal damage is of concern during BLES breast biopsy, the incidence is low, and the outcome of the histopathologic assessment is not affected even in severely damaged specimens. Increased thermal damage seems to correlate with higher fat cell content of the specimen.
METHODS: In this study we included 265 patients with suspicious non-palpable mammographic lesions who underwent 273 stereotactic VABB procedures using the BLES. All biopsies were performed by the same surgical-radiology team under local anesthesia. The specimen thermal damage was classified according to the pathology report in 3 categories: Severe (extensive thermal damage or inability to diagnose), medium (ability to diagnose but either circumferential damage >1.5 mm or diffuse areas of thermal damage) and mild (circumferential thermal damage <1.5 mm but >0.5 mm).
RESULTS: Radiofrequency-associated thermal damage of the specimen was observed in 14 cases (5.13 %), and was classified as severe in 5, medium in 5, and mild in 4 specimens. Within the group of RF damaged specimens, we found a significant (p < 0.05) positive correlation between fat cell content and classification of thermal damage.
CONCLUSIONS: Although thermal damage is of concern during BLES breast biopsy, the incidence is low, and the outcome of the histopathologic assessment is not affected even in severely damaged specimens. Increased thermal damage seems to correlate with higher fat cell content of the specimen.