Investigation of the Reproducibility of Portable Optical Coherence Tomography in Diabetic Macular Edema.

Chiku Yoshiaki, Hirano Takao, Nakamura Marie, Takahashi Yoshiaki, Miyasaka Hideki, Hoshiyama Ken, Murata Toshinori

СтатьяПеревод пока не подготовлен

Аннотация

Background

Diabetic macular edema (DME) causes vision impairment and significant vision loss. Portable optical coherence tomography (OCT) has the potential to enhance the accessibility and frequency of DME screening, facilitating early diagnosis and continuous monitoring. This study aimed to evaluate the reliability of a portable OCT device (ACT100) in assessing DME compared with a traditional stationary OCT device (Cirrus 5000 HD-OCT plus).

Methods

This prospective clinical investigation included 40 eyes of 33 patients with DME. Participants with significant refractive errors (myopia > -6.0 diopters or hyperopia > +3.0 diopters), vitreous hemorrhage, tractional retinal detachment, or other ocular diseases affecting imaging were excluded. Spectral-domain OCT was performed by a single examiner using both devices to capture macular volume scans under mydriasis. Central macular thickness (CMT) was evaluated using the analysis software for each device: Cirrus used version 6.0.4, and ACT100 used version V20. We analyzed inter-evaluator and inter-instrument agreements for qualitative assessments of the intraretinal fluid (IRF), subretinal fluid (SRF), and epiretinal membrane (ERM) using Cohen's kappa coefficient, whereas quantitative CMT assessments were correlated using Spearman's correlation coefficient.

Results

Substantial inter-evaluator agreement for IRF/SRF (κ = 0.801) and ERM (κ = 0.688) with ACT100 and inter-instrument agreement (κ = 0.756 for IRF/SRF, κ = 0.684 for ERM) were observed. CMT values measured using ACT100 were on average 29.6 μm lower than that of Cirrus (285.8 ± 56.6 vs. 315.4 ± 84.7 μm, p < 0.0001) but showed a strong correlation (R = 0.76, p < 0.0001).

Conclusions

ACT100 portable OCT demonstrated high reliability for DME evaluations, comparable to that of stationary systems.

Для цитирования: Chiku Yoshiaki, Hirano Takao, Nakamura Marie, Takahashi Yoshiaki, Miyasaka Hideki, Hoshiyama Ken, Murata Toshinori. Investigation of the Reproducibility of Portable Optical Coherence Tomography in Diabetic Macular Edema.. Pharmaceuticals (Basel, Switzerland). 2024. DOI: 10.3390/ph17101357.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.