Interventional study of the iSelfie heart health tool for assessment of vital signs (heart rate, blood pressure, and Oxygen Saturation) in the Kingdom of Saudi Arabia.

Alhomod Abdulaziz S, Abdulaziz Almowanes Demah, Scooter Plowman Robert, Quddus Azhar, Mohamed Hassan, Karakaya Mahmut, Elgebaly Ahmed, Alrahimi Jamilah, Alkutshan Raed, Tash Adel

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To validate the diagnostic performance of the iSelfie Heart Health tool compared to standard-of-care devices, focusing on accuracy, reliability, and classification capabilities.

METHODS

Vital sign measurements from 579 participants were analyzed using the iSelfie tool in calibrated and non-calibrated modes. Further analyses applied the Association for the Advancement of Medical Instrumentation (AAMI) standards and criteria for cuff-based blood pressure (BP) devices. Primary endpoints included mean absolute error (MAE) and cumulative error percentages for all vital parameters. Binary and multi-class classification analyses evaluated the device's performance to categorize hypertension grades.

RESULTS

In calibrated mode, the iSelfie tool demonstrated high accuracy, with MAE of 4.42 beats per minute (BPM) for heart rate (HR), 8.94 mmHg for systolic BP (SBP), 7.50 mmHg for diastolic BP (DBP), and 1.16% for oxygen saturation (SaO2). When applying AAMI criteria, MAE was reduced to 4.35 BPM for HR, 5.29 mmHg for SBP, and 3.60 mmHg for DBP. Binary classification analysis to distinguish normotensive from hypertensive states yielded 84.4% sensitivity and 84.7% specificity in calibrated mode after applying AAMI criteria. Multi-class classification accuracy was highest for normotensive cases (Grade 0), with limited representation for higher grades.

CONCLUSION

The iSelfie Heart Health tool demonstrated strong performance in measuring HR and SaO2, with improved BP accuracy following calibration and application of AAMI criteria, highlighting its clinical potential as a contactless, user-friendly tool for cardiovascular health monitoring and screening.

Для цитирования: Alhomod Abdulaziz S, Abdulaziz Almowanes Demah, Scooter Plowman Robert, Quddus Azhar, Mohamed Hassan, Karakaya Mahmut, Elgebaly Ahmed, Alrahimi Jamilah, Alkutshan Raed, Tash Adel. Interventional study of the iSelfie heart health tool for assessment of vital signs (heart rate, blood pressure, and Oxygen Saturation) in the Kingdom of Saudi Arabia.. Saudi medical journal. 2025. DOI: 10.15537/smj.2025.46.12.20250739.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.