DOI 10.1097/j.jcrs.0000000000000252

Медицинская библиотека · Journal of cataract and refractive surgery · 2020

Effect of photorefractive keratectomy on agreement of anterior segment variables obtained by a swept-source biometer vs a Scheimpflug-based tomographer.

Salouti Ramin, Kamalipour Alireza, Masihpour Nasrin, Zamani Mohammad, Ghoreyshi Maryam, Salouti Kia, Nowroozzadeh M Hossein

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate agreement of anterior segment variables between Pentacam-AXL and IOLMaster 700 before vs after photorefractive keratectomy (PRK).

SETTING

Salouti Eye Clinic, Shiraz, Iran.

DESIGN

Prospective cohort with interdevice agreement analysis.

METHODS

This study included healthy PRK candidates who were assessed with both devices preoperatively and 6 months after PRK. Only data from the right eye of each patient was analyzed. Pentacam-AXL average keratometry (AvgK) and zonal keratometry in the central 2.5 mm zone (zonal-K2.5) were each compared with mean keratometry (Km) from the IOLMaster 700. Other main outcome measures included vector analysis of corneal astigmatism (J0 and J45), central corneal thickness (CCT), anterior chamber depth (ACD), and white-to-white (WTW) distance. Axial length (AL) measurements by the same devices on a new cohort of 40 patients who had undergone PRK were also assessed. A paired t test was used to assess the interdevice measurement differences, and Bland-Altman analysis was used to calculate the 95% limits of agreement (LoA).

RESULTS

This study included 97 patients. Preoperative vs post-PRK 95% LoAs between Pentacam-AXL and IOLMaster 700 were as follows: AvgK/Km (-0.42, 0.08 diopter [D]) vs (-0.49, 0.18 D); zonal-K2.5/Km (-0.40, 0.32 D) vs (-0.57, 0.74 D); J0 (-0.33, 0.18 D) vs (-0.28, 0.35 D); J45 (-0.28, 0.23 D) vs (-0.24, 0.27 D); pupil pachymetry/CCT (-18, 12 μm) vs (-2.6, 19.6 μm); apical pachymetry/CCT (-17.4, 12.8 μm) vs (-1.7, 20.9 μm); ACD (-0.03, 0.13 mm) vs (-0.03, 0.13 mm); WTW (-0.68, 0.23 mm) vs (-0.63, 0.14 mm); and AL (-0.07, 0.01 mm) vs (-0.07, 0.03 mm), respectively.

CONCLUSIONS

PRK showed a negative impact on interdevice agreement for CCT and corneal power measurements, whereas it did not have a significant effect on the agreement of devices for ACD, WTW, AL, and the J45 astigmatism vectoral component. For IOL power measurement in post-PRK eyes, the 2 devices could be regarded as interchangeable for measuring AL and ACD but not for keratometry readings.

Для цитирования: Salouti Ramin, Kamalipour Alireza, Masihpour Nasrin, Zamani Mohammad, Ghoreyshi Maryam, Salouti Kia, Nowroozzadeh M Hossein. Effect of photorefractive keratectomy on agreement of anterior segment variables obtained by a swept-source biometer vs a Scheimpflug-based tomographer.. Journal of cataract and refractive surgery. 2020. DOI: 10.1097/j.jcrs.0000000000000252.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.