DOI 10.2147/opth.s244079

Медицинская библиотека · Clinical ophthalmology (Auckland, N.Z.) · 2020

Topography-Guided Refractive Astigmatism Outcomes: Predictions Comparing Three Different Programming Methods.

Stulting R Doyle, Durrie Daniel S, Potvin Richard J, Linn Steve H, Krueger Ronald R, Lobanoff Mark C, Moshirfar Majid, Motwani Manoj V, Lindquist Timothy P, Stonecipher Karl G

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To identify the laser programming strategy that will achieve optimal refractive outcomes of LASIK with a topography-guided laser for eyes with a disparity between cylinder measured by manifest refraction and cylinder measured by topography.

SETTING

Six surgeons at 5 clinical sites in the USA.

DESIGN

Retrospective data review.

METHODS

Preoperative, treatment, and postoperative data on 52 eyes that underwent topography-guided LASIK with the WaveLight EX500 Contoura® Vision excimer laser ablation profile in which the vectors representing the preoperative refractive cylinder and the cylinder measured by the WaveLight® Topolyzer™ VARIO Diagnostic Device (Vario cylinder) differed by >/= 0.50D and/or >/= 10 degrees of orientation were analyzed retrospectively. Data were contributed by six surgeons using the laser at 5 different clinical sites. Vector analysis of postoperative cylindrical refractive error and the actual laser programming strategy was used to calculate the cylindrical correction that would, theoretically, have completely eliminated postoperative refractive cylinder. This was compared to expected results using the preoperative manifest cylinder, the topographic cylinder, and the Phorcides Analytic Engine (Phorcides LLC, North Oaks MN; Phorcides). For analysis, subjects were stratified on the basis of the vector difference between Manifest and Topo cylinder (High, >0.75 D; and Low, ≤0.75 D).

RESULTS

The poorest calculated theoretical outcomes were obtained with the manifest refraction (centroid: -0.43, 0.22; mean calculated error vector: 0.56 ± 0.42 D; p=ns). Better outcomes were obtained with the topographically measured refraction (centroid: 0.37, 0.02; mean calculated error vector: 0.47 ± 0.33 D; p=ns). The best outcomes were obtained with Phorcides (centroid: -0.15, 0.06; mean calculated error vector: 0.39 ± 0.28 D; p=ns). The mean error vector magnitude in the Phorcides Low group was significantly lower than for the Manifest and Topo Low groups (0.26 D vs 0.48 D and 0.33 D; p<0.01). The mean error magnitude in the Phorcides High group was nearly 0.25 D lower than for the Manifest High group (0.48 D vs 0.70 D; p<0.01), but was the same as for the Topo High group (0.48 D vs 0.48 D).

CONCLUSION

Our study suggests that using the topographically measured cylinder or the cylinder selected by Phorcides will produce more desirable refractive outcomes than entry of the preoperative refractive cylinder as the basis for correction of myopia and myopic astigmatism with the WaveLight Contoura Vision excimer laser.

Для цитирования: Stulting R Doyle, Durrie Daniel S, Potvin Richard J, Linn Steve H, Krueger Ronald R, Lobanoff Mark C, Moshirfar Majid, Motwani Manoj V, Lindquist Timothy P, Stonecipher Karl G. Topography-Guided Refractive Astigmatism Outcomes: Predictions Comparing Three Different Programming Methods.. Clinical ophthalmology (Auckland, N.Z.). 2020. DOI: 10.2147/opth.s244079.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.