DOI 10.1097/j.jcrs.0000000000000136

Медицинская библиотека · Journal of cataract and refractive surgery · 2020

Evaluation of total corneal power measurements with a new optical biometer.

Shajari Mehdi, Sonntag Ruven, Ramsauer Michaela, Kreutzer Thomas, Vounotrypidis Efstathios, Kohnen Thomas, Priglinger Siegfried, Mayer Wolfgang J

Контролируемое клиническое исследованиеСтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the repeatability and validity of total corneal power measurements (total keratometry [TK]) obtained with a recently introduced optical biometer (IOLMaster 700, Carl Zeiss Meditec AG) to a Scheimpflug device (Pentacam, Oculus).

SETTING

Department of Ophthalmology, Ludwig-Maximilians-University, Munich, Germany.

DESIGN

Prospective randomized controlled trial.

METHODS

The inclusion criteria were a corneal astigmatism of 3.00 diopters (D) or less, no previous ocular surgery, no known corneal irregularities, and no known dry eyes. All eyes were measured 3 times using the optical biometer and the Scheimpflug device. The results were statistically compared using Bland-Altman, within-subject SD, and astigmatism vector analysis.

RESULTS

Ninety-three eyes of 93 subjects were included. Overall, the repeatability of all eyes measured was 0.42 for standard K, 0.40 for TK, 0.45 for total corneal refractive power (TCRP), 0.43 for true net power (TNP), and 0.39 for simulated K. Bland-Altman analysis showed no significant difference between the optical biometer's TK compared with standard K and the Scheimpflug device's simulated K, TNP, and TCRP, the P value exceeding .05 in each case. A difference of >0.50 diopters between astigmatism measurements between TK and TCRP was found in 10 cases compared with standard K and TNP with 0 and 2 cases.

CONCLUSIONS

Repeatability of both devices was high, although measurements of meridian showed a great variability, suggesting that numerous measurements are needed to enhance accuracy. TCRP measurements between recently introduced TK and TCRP cannot be used interchangeably. Future studies are necessary to evaluate which measurement will result in a better outcome when respective measurements are used for toric intraocular lens calculations.

Для цитирования: Shajari Mehdi, Sonntag Ruven, Ramsauer Michaela, Kreutzer Thomas, Vounotrypidis Efstathios, Kohnen Thomas, Priglinger Siegfried, Mayer Wolfgang J. Evaluation of total corneal power measurements with a new optical biometer.. Journal of cataract and refractive surgery. 2020. DOI: 10.1097/j.jcrs.0000000000000136.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.