PURPOSE: To assess the influence of capsular tension ring (CTR) on surgical outcomes of toric and multifocal intraocular lenses (IOLs) in eyes at high risk of zonular instability.

METHODS

Fifty-five eyes of 43 patients who had undergone phacoemulsification and IOL implantation were included in the analysis. They had some risk of zonular weakness, such as pseudoexfoliation, shallow anterior chamber, high myopia, and phacodonesis, or were judged to have unstable zonules during surgery. Toric IOL was implanted in 9 eyes with CTR and 22 eyes without CTR, while multifocal IOL was used in 9 eyes with CTR and 15 eyes without CTR. Manifest refraction, refractive astigmatism, visual acuity, and degree of IOL decentration and tilt measured using swept-source anterior segment optical coherence tomography were analyzed. Axis misalignment of toric IOLs were also evaluated.

RESULTS

In toric IOLs, co-implantation of CTR significantly reduced decentration and axis misalignment of IOL, resulting in better uncorrected and corrected visual acuity after surgery. In multifocal IOLs, combined use of CTR significantly prevented IOL tilt, leading to better intermediate visual acuity. Spherical equivalent and residual astigmatism were not significantly affected by the use of CTR.

CONCLUSIONS

CTR reduces decentration and axis misalignment of toric IOL and tilt of multifocal IOL, achieving improvement of postoperative visual function in eyes with suspected zonular instability.

Для цитирования: Miyoshi Teruyuki, Fujie Shuhei, Yoshida Hironori, Iwamoto Hiroshi, Tsukamoto Hideki, Oshika Tetsuro. Effects of capsular tension ring on surgical outcomes of premium intraocular lens in patients with suspected zonular weakness.. PloS one. 2020. DOI: 10.1371/journal.pone.0228999.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.