MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1038/s41433-019-0744-0

Медицинская библиотека · Eye (London, England) · 2019

Efficacy of toric intraocular lens implantation with high corneal astigmatism within the United Kingdom's National Health Service.

Swampillai Andrew J, Khanan Kaabneh Ali, Habib Nabil E, Hamer Catriona, Buckhurst Phillip J

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To determine the efficacy of toric intraocular lens (TIOL) implantation in cataract surgery patients with high levels of pre-operative corneal astigmatism and ocular co-morbidities in a state funded, National Health Service (NHS) hospital.

METHODS

Retrospective cohort study involving consecutive cases of TIOL implantation in cataract surgery with over 3.00DC of pre-operative corneal astigmatism. Subjects were implanted with the Tecnis TIOL (Abbot Medical Optics) with capsular tension ring stabilisation using the Callisto system (Carl Zeiss Meditec). Visual acuity and refraction were assessed at 4-6 weeks post-operatively. Vector analysis was used to calculate the intended refractive correction, surgically induced refractive correction (SIRC), correction ratio (CR), error of magnitude (EM) and error vector (EV).

RESULTS

Sixty-six eyes of forty-seven subjects aged 73.8 ± 11.9 were included. Eyes with ocular co-morbidities included dry age-related macular degeneration (n = 13), amblyopia (n = 7), high myopia (n = 7), glaucoma (n = 6), previous corneal transplantation (n = 2), nanophthalmos (n = 2) and corneal scarring (n = 1). Pre-operative corneal astigmatism was 4.25 ± 1.69DC (range 3.00-12.00), post-operative refractive astigmatism was 1.31 ± 1.05DC (range 0.00-6.50DC) and post-operative unaided visual acuity was 0.25 ± 0.19 LogMAR. Vector analysis demonstrated an SIRC of 4.08 ± 1.39DC, CR = 1.1 ± 0.3, EM -0.4 ± 1.0 and EV of 1.23 ± 0.72.

CONCLUSIONS

The results demonstrate the efficacy of TIOL implantation in patients with high corneal astigmatism and provide strong evidence advocating their use in cataract surgery within a state funded hospital eye service. Refractive astigmatism was significantly lower than the pre-operative corneal astigmatism and a low error vector was achieved relative to the magnitude of correction.

Для цитирования: Swampillai Andrew J, Khanan Kaabneh Ali, Habib Nabil E, Hamer Catriona, Buckhurst Phillip J. Efficacy of toric intraocular lens implantation with high corneal astigmatism within the United Kingdom's National Health Service.. Eye (London, England). 2019. DOI: 10.1038/s41433-019-0744-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.