DOI 10.1007/s00417-019-04529-2

Медицинская библиотека · Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2019

Small incision lenticule extraction (SMILE) lenticule thickness readout compared to change in axial length measurements with the IOLMaster.

Tay Eugene, Bajpai Ram

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare theoretical values from the small incision lenticule extraction (SMILE) lenticule thickness readout with change in axial length measurements taken with the IOLMaster.

METHODS

We prospectively studied 214 eyes from 107 patients undergoing bilateral SMILE surgery for myopia or myopic astigmatism between December 2014 and May 2017 at an ophthalmological practice in Singapore. All eyes were examined pre-operatively and 1 and 3 months post-operatively with the IOLMaster following SMILE surgery. Achieved lenticule thickness was taken as the change in axial length after surgery. A linear mixed-effects model was used to examine changes in axial length, spherical equivalent and acuity over time. The relationships between change in axial length and theoretical lenticule thickness and spherical equivalent were examined with multiple linear regression analyses, and model prediction was assessed with adjusted R2 statistics.

RESULTS

Mean (95% confidence interval [CI]) spherical equivalent pre-operatively was - 5.25 (95% CI - 5.38 to - 5.12) diopters (D), at 1 month was 0.04 (95% CI - 0.09 to 0.17) D (p < 0.001), and at 3 months was - 0.02 (95% CI - 0.15 to 0.11) D (p < 0.001). Mean (95% CI) pre-operative axial length was 27,726 (95% CI 25,595 to 25,857) μm. Post-operative axial length at 1 month was significantly shorter at 25,595 (95% CI 25,464 to 25,726) μm (p < 0.001) with no change thereafter (p = 0.647). Pre-operative mean ± standard deviation (SD) refractive target was 0.24 (± 0.3) D, and mean difference between target and post-operative spherical equivalent at 1 month was 0.20 D (95% CI 0.16 to 0.25 D, p < 0.001). Multiple regression analysis showed that change in axial length at 1 month was, on average, 5% lower than theoretical lenticule thickness, indicating an average difference of 5.4 μm (95% CI 5.2 to 5.6 μm). Preoperative spherical equivalent predicted negative association with change in axial length at 1 month (β = - 14.8, 95% CI - 18.2 to - 11.3, adjusted R2 = 0.457, p < 0.001).

CONCLUSIONS

Calculated lenticule thickness values were less than expected, and post-operative refractive outcomes at 1 month showed a slight under-correction. Further research in this area is needed to validate these findings.

Для цитирования: Tay Eugene, Bajpai Ram. Small incision lenticule extraction (SMILE) lenticule thickness readout compared to change in axial length measurements with the IOLMaster.. Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie. 2019. DOI: 10.1007/s00417-019-04529-2.