MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/1120672119882334

Медицинская библиотека · European journal of ophthalmology · 2019

Comparative evaluation of corneal incision enlargement after intraocular lens delivery of new preloaded and manual implantation systems.

Wang Lixin, Wolfe Patricia, Paliwal Sumit, Chernosky Alexis, Kohnen Thomas

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To determine whether total corneal incision enlargement after implantation of an intraocular lens with a new preloaded delivery system is comparable to a standard-of-care manual delivery system using an in vitro human cadaver eye model, despite having a smaller initial incision size.

METHODS

Human cadaver phakic whole eye globes were used for these studies (n = 16 per group). Each pair of eyes was randomly assigned to a new preloaded delivery system (UltraSertTM) or a manual delivery system (MONARCH® III D). The surgical procedure included creating a 2.2- and 2.4-mm corneal incision for the preloaded and manual delivery systems, respectively, measuring intraocular pressure and pre-implantation corneal incision size, delivering the intraocular lens into the anterior chamber, and measuring the post-implantation incision size.

RESULTS

The final corneal incision enlargement after intraocular lens delivery using the preloaded delivery system was 2.33 ± 0.04 mm, compared to 2.54 ± 0.05 mm after intraocular lens delivery with the manual delivery system. The mean corneal incision enlargement was comparable between the two systems, being 0.13 ± 0.04 mm using the preloaded delivery system and 0.14 ± 0.05 mm using the manual delivery system (p = 0.432).

CONCLUSION

In a human cadaver eye model, the preloaded delivery system demonstrated an intraocular lens delivery performance on cornea incision enlargement was noninferior to the manual, standard-of-care intraocular lens delivery system despite a smaller initial incision size.

TRANSLATIONAL RELEVANCE: Smaller incision sizes for cataract surgery improve patient outcomes via faster visual and wound recovery and decreased risk of complications such as postoperative inflammation and surgically induced astigmatism.

Для цитирования: Wang Lixin, Wolfe Patricia, Paliwal Sumit, Chernosky Alexis, Kohnen Thomas. Comparative evaluation of corneal incision enlargement after intraocular lens delivery of new preloaded and manual implantation systems.. European journal of ophthalmology. 2019. DOI: 10.1177/1120672119882334.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.