MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.injury.2025.112162

Медицинская библиотека · Injury · 2025

Osseointegration for transfemoral amputees: Influence of femur length and implant sizing on bone-implant contact and alignment.

Cao Matthew, Doherty David B, He Hongjia, Vanodia Vinay P, Reeves Kristin, Ismaily Sabir K, Sauer Jonathan J, Han Shuyang

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Clinical data on osseointegration (OI) for limb replacement indicates a concerning increase in mechanical complications after five years post-implantation. Since adequate bone-implant contact and proper implant alignment are critical factors for successful osseointegration, it is essential to identify the factors influencing these outcomes. This study aimed to assess the effects of residual femur length and implant sizing on bone-implant contact and implant alignment.

METHODS

Three-dimensional models of eight cadaveric femora were reconstructed from CT scans, and transfemoral amputations were simulated for each femur at three levels: short (ST: 1/3 L), medium (MD: 1/2 L), and long (LG: 2/3 L). Virtual OI surgeries were then performed using computer-aided design (CAD) models, and implant sizes were recorded. Subsequently, the effect of implant sizing was assessed by adjusting the implant size by ±1 mm. Contact length and implant alignment were compared between the groups using repeated measures analysis of variance with Bonferroni correction.

RESULTS

The contact in the ST group (65.2 ± 7.3 %) was significantly less compared to the MD (96.1 ± 4.1 %, p < 0.001) and LG (96.8 ± 3.2 %, p < 0.001) groups. Upsizing increased contact in the ST group by 7.0 ± 2.6 % (p < 0.001), though it did not reach levels comparable to the MD and LG groups (p > 0.05). Additionally, sagittal implant misalignment was larger in the ST group (10.1 ± 2.0°) than in the MD (7.9 ± 1.5°) and LG (3.5 ± 1.6°) groups (p < 0.001), while no difference was observed in coronal implant alignment.

CONCLUSION

These findings highlight the factors influencing mechanical complications of osseointegration implants and provide a basis for refined implant designs and surgical techniques for patients with short residual limbs.

Для цитирования: Cao Matthew, Doherty David B, He Hongjia, Vanodia Vinay P, Reeves Kristin, Ismaily Sabir K, Sauer Jonathan J, Han Shuyang. Osseointegration for transfemoral amputees: Influence of femur length and implant sizing on bone-implant contact and alignment.. Injury. 2025. DOI: 10.1016/j.injury.2025.112162.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.