Retinal image quality after microincision intraocular lens implantation.

Alió Jorge L, Schimchak Patricia, Montés-Micó Robert, Galal Ahmed

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the modulation-transfer function (MTF) in eyes implanted with a conventional intraocular lens (IOL) and 2 IOLs designed for microincision cataract surgery (MICS).

SETTING

Research, Development, and Innovation Department, Instituto Oftalmológico de Alicante, Alicante, Spain.

METHODS

This prospective nonrandomized consecutive series comprised 30 eyes implanted with 1 of the following IOLs: conventional acrylic foldable (AcrySof MA60BM, Alcon Laboratories) or the UltraChoice 1.0 ThinOptX (ThinOptX Inc.) or the Acri. Smart 48S (Acri.Tech) MICS IOL. The 0.5 MTF and 0.1 MTF following MICS were calculated 3 months after implantation with the Optical Quality Analysis System (OQAS) for a 5 mm pupil. The differences were statistically analyzed with the Mann-Whitney U test.

RESULTS

The values of 0.5 MTF for AcrySof, UltraChoice, ThinOptX, and Acri. Smart IOLs were, respectively, 2.647 cycles per degree (cpd) +/- 0.833 (SD), 2.601 +/- 0.986 cpd, and 3.453 +/- 0.778 cpd. The mean 0.1 MTF values for the same IOLs were 8.720 +/- 3.074 cpd, 8.814 +/- 4.380 cpd, and 11.418 +/- 2.574 cpd, respectively. Statistical analysis did not show significant differences in 0.5 MTF and 0.1 MTF between the conventional IOL and MICS IOLs.

CONCLUSIONS

Microincision cataract surgery IOLs showed excellent MTF performance when implanted after cataract surgery, equal to that of conventional IOLs. An in vivo MTF study may be an excellent option to evaluate IOL performance in the eye.

Для цитирования: Alió Jorge L, Schimchak Patricia, Montés-Micó Robert, Galal Ahmed. Retinal image quality after microincision intraocular lens implantation.. Journal of cataract and refractive surgery. 2005. DOI: 10.1016/j.jcrs.2005.01.026.