MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2004.06.050

Медицинская библиотека · Journal of cataract and refractive surgery · 2005

Intraocular lens-capsular bag imaging with ultrahigh-resolution optical coherence tomography Pseudophakic human autopsy eyes.

Linnola Reijo J, Findl Oliver, Hermann Boris, Sattmann Harald, Unterhuber Angelika, Happonen Risto-Pekka, Drexler Wolfgang

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare in vitro ultrahigh-resolution optical coherence tomography (UHR OCT) cross-sectional images of the pseudophakic human autopsy eye with histology to evaluate the potential of this imaging technique for enhanced visualization of the anterior segment, especially the capsular bag, intraocular lens (IOL), and posterior capsule opacification (PCO) formation after cataract surgery.

SETTING

Department of Medical Physics and Department of Ophthalmology, University of Vienna, Vienna, Austria, and Department of Oral and Maxillofacial Surgery Institute of Dentistry, University of Turku, Turku, Finland.

METHODS

Ultrahigh-resolution OCT images were acquired from 7 pseudophakic human autopsy eyes using 1.4 microm axial x 3.0 microm transverse resolution. The axial resolution with UHR OCT is 1.4 microm compared to 10.0 microm with the commercially available OCT. Plastic-embedded histologic sections were obtained in precise alignment with the OCT tomograms.

RESULTS

Ultrahigh-resolution OCT cross-sectional tomograms corresponded to the histologic sections. With the wavelength used (800 nm), the anterior and posterior capsules, area of lens epithelial cell growth and extracellular matrix proliferation, and IOL could be clearly visualized. The extent of capsular bag adhesion to the IOL could be detected, as well as the amount of PCO formation.

CONCLUSIONS

The improved resolution makes UHR OCT a powerful tool in anterior segment imaging and evaluation of the capacity of IOL materials and models to induce capsular bag adhesion. Ultrahigh-resolution OCT may also help in determining the area of origin of PCO after cataract surgery.

Для цитирования: Linnola Reijo J, Findl Oliver, Hermann Boris, Sattmann Harald, Unterhuber Angelika, Happonen Risto-Pekka, Drexler Wolfgang. Intraocular lens-capsular bag imaging with ultrahigh-resolution optical coherence tomography Pseudophakic human autopsy eyes.. Journal of cataract and refractive surgery. 2005. DOI: 10.1016/j.jcrs.2004.06.050.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.